Assembly and connection
IMPORTANT!
(a)
(b)
(c)
22

4.4 Vacuum connection

WARNING
Risk of bursting
Prevent uncontrolled overpressure, e. g., when
>
>
connecting to a locked or blocked tubing system.
> Maximum admissable pressure at vacuum sensor:
1,5 bar/1126 Torr (abs.).
> Pollution and damages, especially at the flange, could affect
the measurement.
Connection options
Connection via hose nozzle
Required connection material: Hose nozzle DN 6/10 mm, kurled
nut M14x1, locking ring; optionally: Vacuum hose and compatible
hose clamp (tool: fork wrench size 17).
1. If installed, unscrew the small flange from the vacuum con‑
nection of the gauge.
2. Connect hose nozzle (a), kurled nut (b) and locking ring (c)
as shown in the figure.
Connection via PTFE hose DN 8/10
or
Connection via hose nozzle DN 6/10
or
Connection via small flange KF DN16