CLEANING THE BRACKET SYSTEM (CONTINUED)

CLEANING SOLUTIONS

Phenolic type or quaternary type disinfectants can be used (excluding Virex® TB). Iodophor type disinfectants, Peracetic
Acid, Chlorine, and especially Hydrogen Peroxide are not recommended for use because staining and corrosion may
result.
Suggested cleaners for the Bracket:
9 Quaternary Cleaners (active ingredient - ammonium chloride)
9 Phenolic Cleaners (active ingredient - o-phenyl phenol)
9 Chlorinated Bleach Solution (3.25% - less than 1-part bleach to 100 parts water)
WARNING
9 Some cleaning products are corrosive in nature and may cause damage to the product if used improperly.
9 If the products described are used to clean the equipment, the Bracket and components must be rinsed with clean
water and thoroughly dried following cleaning. Failure to properly rinse and dry the system will leave a corrosive
residue on the surface of the product, possibly causing premature corrosion of critical components.

REMOVAL OF IODINE COMPOUNDS

Use a solution of 0.13 oz / 3.70 ml of Sodium Thiosulfate in 1 pt / 0.5 L of warm water to clean the stained area. Clean
as soon as possible after staining occurs. If stains are not immediately removed, allow solution to soak or stand on the
surface. Thoroughly rinse surfaces with clean water and allow to air dry before returning unit to service.
WARNING
9 Failure to properly clean contaminated parts or components will increase the risk of exposure to blood borne
pathogens and may cause injury to the patient or the operator.

PREVENTIVE MAINTENANCE

A preventive and regular maintenance program should be established for all Technimount equipment. Preventive
maintenance may be required more frequently based on the usage level of the product. The Bracket requires regular
preventative maintenance. Establish and follow a preventive maintenance schedule and keep records of preventive
maintenance activities (see maintenance log in this user guide).
CAUTION
9 Close attention should be given to safety features including, but not limited to locking mechanisms and Discs.
9 Improper maintenance can cause injury or damage to the product. Maintain the product as described in this user
guide.

LUBRICATION

The Bracket System has been designed to operate without the need for lubrication. If you have questions, please
contact Technical Support at techsupport@technimount.com.
CAUTION
9 Failure to use authorized parts, lubricants, etc. could cause damage to the cot/stretcher and will void the warranty
of the product.
Return To Table of Contents
22
Maintenance Guide