Dose Feeding
™
Kangaroo
Connect Enteral Feeding Pump
"Dose Feeding" Directions For Use
Battery Operation
Unplugging the pump from the AC line will automatically put the pump on battery power.
To recharge the battery, plug the pump into an AC wall outlet. The pump will automatically begin
charging.
General Instructions
1. Fill or spike feeding container prior to setting up pump.
2. Turn pump on - Press
Power
button in lower right-hand corner.
3.
In order to achieve proper accuracy, the fluid line in the feeding set bag must be 10 inches above the
top of the feeding pump, when initiating the feeding cycle.
4. Select u
"Keep Settings"
or u
"Clear Settings"
the u
"Keep Settings"
option should be selected).
5. Load the cassette portion of the feeding set into the pump, following the animated instructions
on the screen. Insert from the bottom and snap into place.
Priming the Pump
1. Press u
"Prime Pump"
to access the pump priming options.
2. Press u
"Auto Prime"
to automatically prime the pump set. The pump will quickly prime the feed line
and stop the formula before reaching the end of the feeding line.
3. Press the u
"Hold To Prime"
menu selections to manually top off the line.
before the end of the line. This will prevent formula building up in the moat.
4. Press u "Done".
Setting Dose Feeding
1. Access
"BIOTECH MODE"
by turning off the pump;
the kangaroo is moving, press and release the
2. The screen will change to purple, with the title
3. Select
u"View
Options". Select u"Dose". Select u"Back". Select
4. To return to normal pump operation, restart the pump in the normal manner.
If "Dose" is not available in "More Options", contact your Medical Provider for further information.
Pump sets should not be reused after 24 hours of initial usage.
(for day to day use with the same feeding regimen
If using ENFit
TM
*, stop prime
turning
on the pump; and while the animation of
uupper left
button.
"BIOTECH
MODE".
u"Restart
Pump".
Troubleshooting Guide
Soft Keys
Press a key to select the
option that appears next
to it on the screens
Power Button
Press ONCE to power on.
Press and HOLD to power off
Symptom
Probable Cause
Feed
Dose feeding meets the
Complete
specified amount.
Feed
Dose feeding is interrupted.
Incomplete
This will appear if the pump
Pump
has been without input for
Inactive
more than 10 minutes.
There is an unusual amount
Rotor
of resistance to the rotation
Stuck
of the rotor.
Cassette
Appears if the cassette is not
Dislodged Or
properly loaded in the pump
Cassette Error
set loading area.
Appears if there is a blockage,
Supply Tube
obstruction, or kinked tubing
Blocked
between the feeding bag and
the pump.
Appears when the bag is empty or
Feed Bag
large amounts of foam or bubbles
Empty
are in the feeding solution
15 Hampshire Street Mansfield, MA 02048
ENFit
TM
* is a federally registered trademark in multiple jurisdictions through the world. Used with permission.
© 2019 Cardinal Health Canada. All Rights Reserved. CARDINAL HEALTH, the Cardinal Health LOGO, Kangaroo, and ESSENTIAL TO CARE are trademarks or registered
trademarks of Cardinal Health. All other marks are the property of their respective owners. CA Lit. No. CA-PMR-0010, 18-123-D (2/2019)
Pump Status
Red=Warning Alarm
Yellow=Notice or Caution Alarm
Green=Standby or Feeding
Power Source
Shines a green light when pump
is connected to AC power
Corrections
Press u
DONE
to clear the notification.
Press u
Resume Dose
to continue feeding from the
current pump state or press u
Restart Dose
if you are
beginning a new, full feeding regimen.
Press u
Continue
to return to the previous screen. If the
error cannot be resolved, press and
HOLD
the u
Button
for 3 seconds to stop operation of the pump
and put a different pump into service.
If cleaning of the rotor and rotor shaft does not resolve
this error, press and
HOLD
the u
Power Button
for
3 seconds to stop operation of the pump and put a
different pump into service.
Reload the cassette to ensure the correct positioning.
If the error cannot be resolved, load a new pump set,
prime it, and press u
Continue
to restart the feeding. If
the error cannot be resolved, press and
HOLD
the u
Power Button
for 3 seconds to stop operation of
the pump and put a different pump into service.
Check the tubing between the bag and the pump. If the
tube cannot be cleared, remove the blocked pump set
and load a new pump set onto the pump. If the error still
cannot be resolved, press and
HOLD
the u
Power Button
for 3 seconds to stop operation of the pump and put a
different pump into service.
If the bag is empty, re-fill the bag as required. If the bag still
contains feeding solution, remove cassette and check the
bag side tubing for excessive foam or bubbles. Clear bubbles
from line and reload the feeding set or replace with a new
feeding set. If the error still cannot be resolved, press and
HOLD
the u
Power Button
for 3 seconds to stop operation of
the pump and put a different pump into service.
Power
Need help?
Do you have a question about the Kangaroo Connect and is the answer not in the manual?
Questions and answers