Quick Start
™
Kangaroo
Connect Enteral Feeding Pump
"Quick Start" Directions For Use
IMPORTANT:
•
Before powering on, please ensure that no cassette is inserted in the pump, this could trigger an error message.
•
Wait 3 minutes after pouring formula in bag before priming and running the pump.
•
Wipe down the inside of the downstream sensor with a clean damp cloth on a daily basis.
Pump Overview:
1
2
3
Setting up the pump:
1. Press the
Power
button.
2. Load set (insert from the bottom and snap into place).
3. Select Prime.
1
Priming the pump:
1. Select
Prime
mode.
2. If you selected
Auto
Prime, wait for pump to finish priming.
3. If you selected
Hold to
Prime, press and hold
1
1.
Soft Keys
- Press a key to select the option
that appears next to it on the screen.
2.
Power Button
- Press once to power on.
4
Press and hold to power off.
3.
Power Source
- Shines a green light when
the pump is connected to AC power.
Pump Status
4.
= Warning Alarm
= Notice or Caution Alarm
= Standby or Feeding
2
Prime Feed
until fluid reaches end of tubing.
2
3
Starting a feeding:
1. Select
Clear
Settings.
2. Use the
plus (+)
and
minus (-)
3. When ready to begin feeding, select Start.
4. To suspend the feeding, select Pause.
5. To resume feeding, select Continue.
1
4
Powering down the pump:
Press and hold the
Power
button for 3 seconds.
3
Ensuring the correct fluid level:
Make sure the top of the fluid column is placed at 10 inches (25.4 cm) above the center of the pump rotor.
10 in. (25.4 cm)
15 Hampshire Street Mansfield, MA 02048
© 2019 Cardinal Health Canada. All Rights Reserved. CARDINAL HEALTH, the Cardinal Health LOGO, Kangaroo, and ESSENTIAL TO CARE are trademarks or registered
trademarks of Cardinal Health. All other marks are the property of their respective owners. CA Lit. No. CA-PMR-0008, 18-123-E (2/2019)
buttons to enter the rate, then select OK.