This manual describes the installation and use of the two models in the U1000MkII range: • U1000MkII-FM is an ultrasonic clamp-on flow meter for measuring flow rate and total flow with a volume pulse output and optional Modbus, M-Bus or 4-20mA flow proportional output. It can be used as a standalone meter or as part of an integral management system.
U1000MKII User Manual Micronics How Does It Work? The U1000MKII uses a cross correlation transit time algorithm to provide accurate flow measurements. An ultrasonic beam of a given frequency is generated by applying a repetitive voltage pulse to the transducer crystals. This transmission goes first from the downstream transducer to the upstream transducer as shown in the upper half of Figure 1.
Micronics U1000MKII User Manual Package Contents The unit consists of two parts: Electronics Module Consisting of the keypad and display, power, signal and comms connections. The Electronics Module clips onto the Sensor Assembly. Sensor Assembly Incorporating guide rails and two transducers for flow measurement.
U1000MKII User Manual Micronics Display Flow Signal strength Flow Units Flow Direction Measured flow. For accurate Units applicable to * Indicates positive results, must be flow measured flow. over 40%. ! Indicates reversed flow. 4-20mA (1.) 4-20mA function is Enabled.
Important: Do not expect to obtain accurate results if the unit is positioned close to any obstruction that distorts the uniformity of the fluid flow profile. Micronics ltd accepts no responsibility or liability if product has not been installed in accordance with these instructions.
U1000MKII User Manual Micronics 2.1.2 Clean the Pipe's Flow Sensor Contact Area Prepare the pipe by removing any paint and degreasing the area to be installed to obtain the best possible surface. A smooth contact between pipe surface and the face of the sensors is an...
Micronics U1000MKII User Manual 2.2.2 Pulse Output Connection The isolated pulse output is provided by a SPNO/SPNC MOSFET relay which has a maximum load current of 500 mA and maximum load voltage of 24 V AC/DC. This output is suitable for SELV circuits only.
The initial screen sequence is different for the FM and HM models. 2.3.1 U1000MkII-FM Switch on the power to the Electronics Module. A Micronics start-up screen is displayed for 5 seconds followed by hardware and software version information. You are then prompted to enter the internal diameter...
Page 13
Micronics U1000MKII User Manual The unit now shows the flow sensor separation Set Separation: Code Make a note of the Code. Is needed during the installation. All subsequent start-ups will use the same configuration. Continue with the installation of the Sensor Assembly (see page 12).
U1000MKII User Manual Micronics 2.3.2 U1000MkII-HM Switch on the power to the Electronics Module. A Micronics start-up screen is displayed for 5 seconds followed by hardware and software version information. You are then prompted to enter the internal diameter Enter Pipe ID: of the pipe: 050.0 mm...
Page 15
Micronics U1000MKII User Manual For Water the Temperature range is entered Temperature: For a temperature of <= 40.0 °C select “COLD”. HOT | COLD For a temperature of > 40.0 °C select “HOT”. The unit now shows the correct flow sensor Set Separation: separation (in this case, “B-2”) for the chosen values...
U1000MKII User Manual Micronics 2.4 Adjust Flow Sensor Separation Using the separation code displayed by the Electronics Module, take the Sensor Assembly and adjust the flow sensor separation accordingly: Adjust the screws on the flow sensors as to allow sideways movement. DO NOT fully unfasten or remove the screws at this stage.
Micronics U1000MKII User Manual 2.6.1 Pipe Adaptors The diagrams below show how the adaptors are fitted. The top ‘V’ shaped adaptor clips onto the ends of the Sensor Assembly and this should be used with all pipes with an outside diameter less than 60 mm.
U1000MKII User Manual Micronics Using the hose clips provided, clamp the Sensor Assembly (and adaptors, if used) to the pipe at an angle of 45° to the top of the pipe. Experience has shown that the most consistently accurate results are achieved when the unit is mounted at this angle. This minimises the effect of any flow turbulence resulting from entrained air along the top of the pipe and sludge at the bottom.
Micronics U1000MKII User Manual 2.8 Connect Electronics Module Ensure that the power is switched off. Connect the Electronics Module. The two leads can be connected either way round. The SMB connections are fragile so be careful when installing not to snap them off.
U1000MKII User Manual Micronics Enter the password-controlled menu and scroll to the Calibration sub-menu. Press the Enter key until the Zero Temp Offset screen is displayed. Select Yes and press the Enter key to display the Attach Sensors screen. Press the Enter key again and wait for instrument to return to the Zero Temp Offset screen.
Page 21
Micronics U1000MKII User Manual Note: The fluid in the pipe must be in a stationary condition, if there is flow running in the pipe and the zero flow is set it will either result in an error if the flow is high enough or it will zero out the current flow value.
U1000MKII User Manual Micronics 3 MENU STRUCTURES The password-protected menus allow you to change the default settings: Accessing the Menus Ensure that the instrument is in Flow Reading, Total Flow, Temperature dT, Total Energy, Instant Power or Total Flow modes, then press Sig:87% 246.3...
Micronics U1000MKII User Manual 3.4 Volume Total Menu Reset Vol. Total: NO | YES Set Total Form: Float User Menu: Volume totals 3.5 Current Output Menu Select 4-20mA: ON | OFF Flow @ 20mA: 1000.0 l/m Flow @ 4mA: 0000.0 l/m...
U1000MKII User Manual Micronics 3.6 Pulse Output Menu NOTE: SCREENS WITHIN THE RED BOX ARE ONLY SHOWN ON U1000MKII-HM MODELS. Select Pulse: ON | OFF Pulse Type: Pulse Type: Pulse Type: Pulse Type: Flow Alarm Energy Volume Frequency Volume per Pulse...
Micronics U1000MKII User Manual 3.7 Calibration Menu Note: Screens within the red box are only shown on U1000mkII-HM models. Damping Time [s]: Damping Time [s]: Damping Time [s]: Zero Cut-off: 0.01 m/s Clear Damping Time [s]: Zero Offset Zero Offset: Averaging…9...
U1000MKII User Manual Micronics 3.8 Diagnostics Menu The diagnostics menu provides some additional information about the flowmeter and its setup. The menu can be accessed by pressing the key from the main flow-reading screen. Press the keys to move between the diagnostics screens.
Micronics U1000MKII User Manual 4 OUTPUTS Pulse Output Pulse output can be set up to operate one of five modes: • Volumetric • Energy (U1000MkII-HM only) • Frequency • Low Flow Alarm • Loss of Flow (Signal) Alarm The Alarm functions allow you to set the alarm switch to Normally Open or Normally Closed.
U1000MKII User Manual Micronics 4.1.3 Energy Pulse (U1000MkII-HM only) When the Pulse Output is set to Energy, the kWh LED will be permanently illuminated. Choose from 1,10,100 kWh or 1 MWh when in metric mode and 1,10,100 kBTU or 1 MBTU in imperial mode.
Micronics U1000MKII User Manual 5 CALIBRATION Where the unit is factory calibration checked for compliance, the installation should also be verified by means of a calibrated reference meter and the unit calibration adapted to match should it be required. Page 27...
U1000MKII User Manual Micronics 6 RELOCATING THE UNIT If it is necessary to relocate the unit use the following procedure: Disconnect the temperature sensors (U1000MkII-HM only) and MODBUS cable (if used). Unfasten hose clips and remove the complete unit from the pipe.
7.6 Positioning For accurate measurements, the U1000MKII-FM/U1000MKII-HM must be installed at a position where the fluid flows uniformly. Flow profile distortions can result from upstream disturbance such as bends, tees, valves, pumps and other similar obstructions. To ensure a uniform flow profile, the unit must be mounted away from any cause of flow disturbance.
Page 34
Important: Do not expect to obtain accurate results if the unit is positioned close to any obstruction that distorts the uniformity of the flow profile. Micronics ltd accepts no responsibility or liability if product has not been installed in accordance with these instructions.
Again this can vary as levels are topped-up. 7.11 Limitations with composite pipes Testing has show that there is limited compatibility with the Micronics ultrasonic flowmeters when being used with a multi-layer composite pipe. The construction of the pipe doesn’t allow signals to propagate in the desired manner therefore the signals can either arrive in an unexpected location or the signal is so diminished that there is no information to process.
U1000MKII User Manual Micronics 8 APPENDIX Specification General Measuring Technique Transit time Measurement channels Timing Resolution ±50 ps Turn down ratio 100:1 Flow velocity range 0.1 to 10 m/s Applicable Fluid types Clean water with < 3% by volume of particulate content, or up to 30% ethylene glycol.
U1000MKII User Manual Micronics 8.2 Default values The settings will be configured at the factory for metric units. The following table lists the metric and imperial default values. Parameter Default Value Metric Imperial Dimensions inches Flow Units l/min USgal/min Pipe size (ID) 1"...
Micronics U1000MKII User Manual 8.2.1 Data Entry Errors These generally advise you that the data entered is not within the specified range (Replace with Imperial units as needed): Displayed when an invalid Pipe ID is entered, prompting the user Range 20.0–165.1 to enter a value between 20 and 165 mm, depending on the 0.000 mm...
Need help?
Do you have a question about the U1000MKII-FM and is the answer not in the manual?
Questions and answers