CAUTION: Cold Environments
If the AED Pro is stored in an environment with a temperature below the operating temperature, the unit should be
allowed to warm up to the needed operating temperature before using.
CAUTION: Not a Patient Monitor
The AED Pro is not a true patient monitor with the requisite alarms. Medical personnel should attend patients at all
times while the AED Pro is in use.
CAUTION: The AED is programmed with software that has been tested to work with versions of ServiceLink and
RescueLink that are included with the AED. When using older versions of ServiceLink and RescueLink to
communicate with this AED, there may be features described in this manual that are not available to be used.
Also, when communicating with an older AED with the version of ServiceLink and RescueLink included with this
new AED there may be features described in this manual that cannot be used. The software in most cases will
give an error message when incompatibilities occur.

SYMBOL DESCRIPTIONS

The following symbols may appear in this manual, on the AED Pro, or on its optional components. Some of the symbols
represent standards and compliances associated with the AED Pro and its use.
Dangerous Voltage: The defibrillator output has high voltage and can present a shock hazard. Please read and
understand all safety alerts in this manual before attempting to operate the AED Pro.
Attention!: Identifies important information in this manual, on the AED Pro, or on its component parts regarding
the safe and proper use of the AED Pro.
Defibrillator Proof Type BF Equipment: The AED Pro, when connected to the patient's chest by the pads, can
withstand the effects of an externally applied defibrillation shock.
CE Mark: This equipment conforms to essential requirements of the Medical Device Directive 93/42/EEC.
The AED Pro is protected against the effects of splashing water in accordance with IEC 60529.
IP24
Classified by ETL Semko with respect to electric shock, fire and mechanical hazards only in accordance with
UL 60601-1, CAN/CSA C22.2 No.601.1-M90, EN60601-1 and EN60601-2-4. Conforms to UL Standard
UL60601-1. Certified to CAN/CSA Standard C22.2 No. 601.1-M90.
International symbol for ON. Open the lid to turn on the AED Pro.
Open the lid to turn ON the AED Pro.
Indicates the AED Pro battery status. The illuminated areas indicate the remaining battery capacity.
0%
100%
Indicates AED Pro requires maintenance by authorized service personnel.
When the SHOCK indicator is lit, push this button to deliver a defibrillation shock.