2 Installing the Bravo Platform

Installing the pipette head

Installing the pipette head
About this topic
About the head mount
26
This topic describes how to mount the pipette head when you first set up the
Bravo Platform. The procedure is for the current Bravo pipette heads,
including the fixed- tip pipette heads, disposable- tip pipette heads (Series III),
pin tools, and the like.
Note: If you want to install a Series II pipette head on a Bravo Platform that
has a gripper, you must ensure that the gripper is lowered from the docked
position before attempting to install the pipette head.
Each Bravo pipette head contains a dovetail interface that attaches to the
dovetail connector on the Bravo head mount. Two head- retainer pins and a
head lock secure the pipette head to the Bravo Platform.
Before mounting a pipette head, familiarize yourself with the features of the
dovetail interface on the pipette head top and the connector underneath the
head mount. See the following figures.
Figure Bravo head mount with (1) head lock and (2) dovetail connector
00218
Bravo
pendant connection
1
2
Bravo Automated Liquid-Handling Platform User Guide