Non-invasive blood pressure
Hide thumbs Also See for M Series:
Table of Contents

Advertisement

Message/Symptom:
REPEAT NIBP MEAS
alternating with NIBP
SIGNAL WEAK
SYSTEM FAULT
False high reading
False low reading
22
Problem:
Weak or no oscillometric signal.
Communication problem with the NIBP
module.
Cuff too small.
Cuff not centered over brachial artery.
Cuff too loose.
Patient's limb below level of heart.
Cuff too large.
Limb above level of heart.
ZOLL M Series Non-Invasive Blood Pressure
User Action:
Check cuff fit and positioning.
Check hose connection to unit.
Check for kinked hose.
Increase cuff inflation pressure if clini-
cally appropriate.
Cycle power and retry.
If problem persists, return for service.
Select larger cuff.
Readjust cuff.
Reapply cuff or select smaller cuff.
Raise patient's limb to heart level.
Select smaller cuff
Lower patient's limb to heart level.