10 Calibration and adjustment

10.1 Hilti Calibration Service

We recommend that the tool is checked by the Hilti Calibration Service at regular intervals in order to verify its reliability
in accordance with standards and legal requirements.
Use can be made of the Hilti Calibration Service at any time, but checking at least once a year is recommended.
The Calibration Service provides confirmation that the tool is in conformance, on the day it is tested, with the
specifications given in the operating instructions.
The tool will be readjusted if deviations from the manufacturer's specification are found.
After checking and adjustment, a calibration sticker applied to the tool and a calibration certificate provide written
verification that the tool operates in accordance with the manufacturer's specification.
Calibration certificates are always required by companies certified according to ISO 900x. Your local Hilti Center or
representative will be pleased to provide further information.

11 Care and maintenance

NOTE
Have damaged parts replaced by Hilti Service.

11.1 Cleaning and drying

Blow any dust off the glass.
CAUTION
Do not touch the glass surfaces with your fingers.
Use only a soft, clean cloth to clean the tool. If necessary,
the cloth may be moistened with a little pure alcohol or
water.
CAUTION
Do not use liquids other than alcohol or water. Other
liquids may damage plastic parts.
NOTE
Have damaged parts replaced.

11.2 Storage

NOTE
Do not put the tool into storage when wet. Allow it to dry
before putting it away.

12 Troubleshooting

Fault
The tool can't be switched on.
E01
TOO FAST
NOTE
Always clean the tool, its transport container and ac-
cessories before putting them into storage.
NOTE
Check the accuracy of the equipment before it is used
after a long period of storage or transportation.
CAUTION
Remove the battery if the tool is to remain unused for a
long period of time. Leaking batteries may damage the
tool.
NOTE
Observe the specified temperature limits when storing
your equipment, above all in winter and summer, es-
pecially if the equipment is stored in a motor vehicle
(-30°C to +70°C (-22°F to +158°F)).

11.3 Transport

CAUTION
The battery must be insulated or removed from the
tool before the tool is shipped or sent by mail. Leaking
batteries may damage the tool.
Use the Hilti shipping box or packaging of equivalent
quality for transporting or shipping your equipment.
Possible cause
No electric power.
Counting error, when the measured
value displayed changes constantly
when an object is targeted.
The telescope was pivoted too
quickly for the vertical sensor.
Remedy
Charge the battery according to in-
structions.
The tool needs to be repaired.
Pivot more slowly.