Remote Control; Troubleshooting - UNI-T MSO Series Quick Start Manual

Table of Contents

Advertisement

Quick Start Guide

Remote Control

9.
MSO/UPO3000E series digital phosphor oscilloscope communicates with the computer through USB
interface and LAN interface to achieve remote control. Remote control is implemented based on the SCPI
(Standard Commands for Programmable Instruments) command set.
MSO/UPO3000E series digital phosphor oscilloscope supports three remote control modes:
a. User-defined programming
b. Use PC software (instrument manager)
c. Web Control

Troubleshooting

10.
(1)
When press the button, the oscilloscope is blank screen.
Check whether the power is correctly connected, power supply is normal or not.
a.
Check whether the power switch is really turned on. The soft power switch button should light in
b.
green when the instrument is normal boot. And it should have relay sound.
If there is a relay sound, it indicates that the oscilloscope starts normally.
c.
Try the following steps: press the DEFAULT key, then press F1, if device returns to normal, it means
backlight brightness is set too low.
After completing the above steps, restart the oscilloscope.
d.
If the instrument still cannot work, please contact UNI-T.
e.
(2) After signal acquisition, signal of waveform does not appear on the screen.
a.
Check whether BNC port is connected properly.
b.
Check whether the signal output channel has turned on.
c.
Check whether the input signal channel has turned on.
d.
Check whether has DC displacement in signal of information source.
e.
Plug out the input signal, check whether the baseline is in the screen range (if not, please perform
auto-calibration.)
f.
If the instrument still cannot work, please contact UNI-T.
(3) The measured voltage amplitude value is 10 times larger or smaller than the actual value: Check the
channel probe attenuation coefficient setting is consistent with the used probe attenuation rate.
(4) There is a waveform display but not stable
a.
Check the trigger source in the trigger menu and confirm that it matches the input channel of the
actual signal.
b.
Check the trigger type: general signal should use "Edge" trigger. The waveform can display stably
only the proper trigger type is used.
c.
Change the trigger coupling setting to high-frequency or low-frequency reject, in order to filter out
the high-frequency or low-frequency noise that interferes with the trigger.
(5) No display after pressing Run/Stop button
Instruments.uni-trend.com
MSO/UPO3000E Series Oscillosc ope
16 / 17

Advertisement

Table of Contents
loading

Table of Contents