8.4 Auto-collector operating procedures

IMPORTANT: The Auto-collector software does not communicate with the PILS and
therefore has no information about liquid flow rate. It is up to the user to configure the
vial fill time properly so that the vials are filled with an appropriate volume of water.
WARNING: When installing the vial-holding carousel to the carousel turn ta-
ble, be sure to properly align the locating pin on the turn table with the small hole in
the vial-holding carousel. Improper installation of the carousel will cause the needle
to strike the carousel resulting in damage and failure of the Auto-collector. See figure
25 for proper orientation.
Ensure that the small locating pin on the carousel turn table is inserted properly inside
the mating hole on the carousel. Note that an arrow cutout on the carousel turn table
points to the locating pin for easy reference. Once the carousel is resting flat on the turn
table, install and tighten the three wing nuts to ensure the carousel is held tightly during
sampling.
Be sure to drain the Auto-collector drain bottle when it is full.

Auto-collector Software

Brechtel provides software to control the Auto-collector. A screenshot of the Auto-collector
software immediately after first running the software is shown in figure 26, this step re-
83-00009-01A
8.4 Auto-collector operating procedures
PILS 4001-115V and 4001-230V
Figure 25: Photo of a
properly installed vial-
holding carousel.
Page 53 of 74