System Setup; Packing List; Assembly - BRECHTEL 4001-115V Instrument Manual

Particle into liquid sampler (pils)
Table of Contents

Advertisement

7

System Setup

7 System Setup

7.1 Packing List

The PILS will arrive in a sturdy box holding the PILS chassis, bottles, cables, any or-
dered accessories, and the ship kit. If accessories were ordered with the PILS they may
be packaged in the in additional boxes.
See the Appendix for a complete list of all the parts that come with the PILS. Table 5
shows the parts needed to prepare the PILS for operation.
Note: The PILS is shipped from the factory with all water removed to avoid the possibility
of damage to the various components due to freezing during shipment. Before running
the instrument control software ("BMI PILS 4001.exe") be sure to fill the water supply
bottles.
IMPORTANT: The inorganic vapor denuders (if purchased as an option) are shipped
from the factory uncoated. It is the customer's responsibility to properly coat the in-
side surfaces of the denuder before first use. See the denuder rejuvenation procedures
in the maintenance section for more information.

7.2 Assembly

The PILS will arrive with the liquid supply and drain bottles uninstalled. Follow the dis-
cussion below together with the figure and table to assemble the various pieces.
83-00009-01A
PILS 4001-115V and 4001-230V
Page 39 of 74