8.3 Description of PILS output file

One file is created during each PILS run cycle, this data file contains recorded tip and
steam heater temperatures from the temperature controllers. The date, time and num-
ber of seconds since the start of the run cycle are included in first, second and third
columns, respectively, of the data file. The next two columns represent the tip tem-
perature and steam heater temperature. In the header of the file various run cycle infor-
mation and descriptors of the data columns are reported. Be sure to account for liquid
sample delay times when comparing PILS-derived chemical composition data to other
data sets that may have different measurement start times and time resolutions.
83-00009-01A
8.3 Description of PILS output file
PILS 4001-115V and 4001-230V
Figure 24: Front panel
display of BMI PILS 4001
v1.0.exe software showing
measured temperatures
and normal operation.