Temperature Controllers - BRECHTEL 4001-115V Instrument Manual

Particle into liquid sampler (pils)
Table of Contents

Advertisement

IMPORTANT: Avoid installing tubes of different inside diameters in the pump chan-
nels controlling the impactor wash flow and the liquid sample outlet flow. Tubing with
the same inside diameter, material and age should be installed in the two channels
controlling the impactor washflow and liquid sample outlet flow to ensure the flow
rates are the same.
Table 1: Selection of peristaltic pump tubing available for the PILS.
Stop Color
ID
(mm)
Orange/White
0.63
Orange/Blue
0.25
Orange/Red
0.19
Orange/Black
0.13
Red/Red
1.14
Grey/Grey
1.29
Note that even though flow rates higher than 1.5 to 2 ml/min are shown in table 1, the
steam heater flow rate must be 1.5 ml/min under normal operating conditions and the
impactor washflow and liquid sample outlet flows must not exceed 2 ml/min under nor-
mal operating conditions. Once the peristaltic pump speed is set to deliver 1.5 ml/min to
the steam heater, different impactor washflow and liquid sample outlet flow rates are
set by installing tubing with different inside diameters - NOT BY CHANGING THE PUMP
SPEED.

Temperature Controllers

The heater temperature controller controls the power delivered to the heater and the
time that the heater is on. The tip temperature is controlled by the steam heater temper-
ature and is only displayed by the temperature controller labelled 'Steam Tip'. A photo
of the controllers mounted in the PILS is shown in figure 2. The 'Steam Tip' temper-
ature controller does not control anything, it is only used to display the tip tempera-
ture. The tip temperature should be kept within the range 97.5-100 C. The user must
manually adjust the temperature setpoint of the heater temperature controller (labelled
83-00009-01A
Used On
Impactor
Low Flow Kit
Low Flow Kit
Low Flow Kit
Steam Heater
Waste lines
PILS 4001-115V and 4001-230V
6.3 PILS Chassis
Max flow
Flow rate
ml/min
ml/rev
2.6
0.024
0.47
0.004
0.30
0.003
0.10
0.001
8.8
0.08
10.0
0.95
Page 24 of 74

Advertisement

Table of Contents
loading

This manual is also suitable for:

4001-230vPils-ac80Pils-ac38

Table of Contents