Too Long Fiber; Fiber Over Angle; Loss Limit Over; Fiber Thin Error - UCL SWIFT SWIFT KF4 User Manual

Optical fiber arc fusion splicer
Table of Contents

Advertisement

Move the cursor downward.
Move each motor to manual mode.
Heater popup menu should load.
Initialize the splice function.
Return to the menu screen.
o long fiber
7.3|

TOO LONG FIBER

Complete a selection.
Follow the next step on the menu screen.
sage generated when the fiber is located too close to the electrodes, object lens or
Error message generated when the fi ber is located too close to the electrodes, object lens or refl ector is dirty,
long fiber
dirty, or the LED is not bright enough.
or the LED is not bright enough.
Splice execution.
age generated when the fiber is located too close to the electrodes, object lens or
SOLUTION:
Press
and reposition fi ber. Clean the object lens and refl ector. Conduct an LED test. If an error
ress
and reposition fiber. Clean the object lens and reflector. Conduct an LED test.
Return to the initial screen and initialize splice function.
dirty, or the LED is not bright enough.
occurs upon performing the LED test, contact UCLSWIFT.
occurs upon performing the LED test, contact UCLSWIFT.
Turn on the heater.
When it is on, the lamp on the left is in the red.
Press once more when it is on, and the heater is turned off.
ress
and reposition fiber. Clean the object lens and reflector. Conduct an LED test.
occurs upon performing the LED test, contact UCLSWIFT.
SWIFT KF4-EN, Rev.C
er over angle
sage generated when the cleaved angle of the fiber is higher than specified.
7.4|

FIBER OVER ANGLE

er over angle
Error message generated when the cleaved angle of the fi ber is higher than specifi ed.
heck the state of the fiber cleaver. Check the cleaved angle limit.
age generated when the cleaved angle of the fiber is higher than specified.
SOLUTION:
Check the state of the fi ber cleaver. Check the cleaved angle limit.
heck the state of the fiber cleaver. Check the cleaved angle limit.
, Rev.C
Rev.C
49

7.5. Loss limit over

Error message generated when the estimated loss value is higher than the preset loss factor limit.
Solution: Check the loss factor limit.
9

7.6. Fiber thin error

Error message generated when the spliced point becomes thinner than the standard after
splicing.
Solution: Make an adjustment to shorten the pulling length of the pulling splice. Check whether the
arc amount or arc time is set as too large or too long.

7.7. Fiber thick error

Error message generated when the spliced point becomes thicker than the standard after
splicing.
Solution: Reduce the overlap set value. Check whether the arc amount or arc time is set as too
small or too short.

7.8. Bubble error

Error message generated when there are bubbles or spots being generated on spliced point
after splicing.
Solution: Examine the fiber cleaver. Clean the V-groove. Examine the electrodes.

7.9. Cleaved surface error

Error message generated when the cut surface of the fiber is of poor quality.
Solution: Check the condition of the fiber cleaver. Re-cleave the fiber.
SWIFT KF4-EN, Rev.C
66
66
7.5|
LOSS LIMIT OVER
Error message generated when the estimated loss value is higher than the preset loss factor limit.
SOLUTION:
Check the loss factor limit.
7.6|
FIBER THIN ERROR
Error message generated when the spliced point becomes thinner than the standard after splicing.
SOLUTION:
Make an adjustment to shorten the pulling length of the pulling splice. Check whether the arc
amount or arc time is set as too large or too long.
7.7|
FIBER THICK ERROR
Error message generated when the spliced point becomes thicker than the standard after splicing.
SOLUTION:
Reduce the overlap set value. Check whether the arc amount or arc time is set as too small or
too short.
7.8|
BUBBLE ERROR
Error message generated when there are bubbles or spots being generated on spliced point after splicing.
SOLUTION:
Examine the fi ber cleaver. Clean the V-groove. Examine the electrodes.
7.9|
CLEAVED SURFACE ERROR
Error message generated when the cut surface of the fi ber is of poor quality.
SOLUTION:
Check the condition of the fi ber cleaver. Re-cleave the fi ber.