1.3 Special operations

1.3 Special operations

1.3.1 Offset calibration

1.3.2 Factory reset

SR446 400 MHz Programmable Gain Preamplifier
The SR446 has an automatic offset calibration feature that will trim the
input offset voltage to within ± 1 mV of zero after operation. To trim the
DC offset:
1. Ensure the unit has been on for at least 30 minutes to warm up.
2. Disconnect cables from the input and both output ports.
3. Select input grounded by pressing the [Couple] button until
is illuminated.
4. With the
indicator illuminated, press and hold the [Couple]
GND
button until the display shows
5. Release the [Couple] button.
The display will show a countdown while each gain configuration is
separately trimmed and recorded to non‑volatile memory. When the
procedure has completed, the display will briefly show
OCAL command (§2.3.8, pg 19).
Restoring the factory default settings of the SR446 will erase all
user settings and return the instrument to the *RST configuration.
Additionally, any trim data from offset calibration (see section 1.3.1,
above) is erased and returned to the original factory calibration. See
the FRST command (§2.3.8, pg 19).
To initiate a factory reset from the front panel:
1. Turn off the unit.
2. Press and hold the [Couple] button, and then press the [POWER]
button while continuing to hold [Couple] depressed.
3. The display will show
4. After about 5 seconds, the unit will return to normal operation.
£
cAL .
¢
¡
£
rSt
¢
¡
11
GND
£
End . See the
¢
¡