Opc Battery; Alerts - Aqua Lung PHODS User Manual

Portable helicopter oxygen delivery system
Hide thumbs Also See for PHODS:
Table of Contents

Advertisement

20

OPC BATTERY

Alerts (Fig. 34)
PRE-MISSION
POWER-UP
REPLACE BATTERIES
IMMEDIATELY
(prior to mission)
REPLACE BATTERIES
IMMEDIATELY
(prior to mission)
REPLACE BATTERIES
IMMEDIATELY
(Inoperable "Lock-Out" Mode)
Notes:
Replace the batteries immediately if any of the LOW BATTERY Alerts are indicated during the
PHODS PRE-FLIGHT CHECKLIST. DO NOT embark on a mission with weak batteries. A set of
fresh AA alkaline batteries should always be included as part of a "spares" kit.
If a mission is begun with a LOW BATTERY-1 Alert, the OPC unit may have no more than 2 hours
of operation remaining.
If a mission is begun with a LOW BATTERY-2 Alarm, the OPC unit may be on the verge of imminent
shut down.
If a LOW BATTERY-2 Alarm is received during a mission, the batteries should be replaced at the
first opportunity.
If operation is continued with a LOW BATTERY-2 Alarm and subsequent FLOW-FAULT Alarms are
received, this likely indicates that the oxygen-dispensing valve has ceased operating, and the OPC
unit is no longer dispensing oxygen (see BATTERY LIFE AND DEPLETION).
It is particularly important to heed BATTERY ALERTS in the course of missions conducted at higher altitudes.
The times for remaining operation ("Time-to-Failure") are approximations only, based on nominal
conditions (see BATTERY LIFE AND DEPLETION).
OPC
(Nominal)
TEST
LOW BATTERY-1
LOW BATTERY-2
BAD BATTERY
(Nominal)
LOW BATTERY-1
LOW BATTERY-2
LOW BATTERY-2
with FLOW-FAULT
Figure 34
PHODS User's Manual
IN-MISSION
less than 8 hrs remaining
REPLACE BATTERIES
AT FIRST OPPORTUNITY
less than 2 hrs remaining
REPLACE BATTERIES
AT FIRST OPPORTUNITY
No Longer Dispensing Oxygen
OXYGEN SYSTEM SHUTDOWN

Advertisement

Table of Contents
loading

This manual is also suitable for:

102935m2

Table of Contents