Tektronix AWG410 Service Manual page 207

200 ms/s arbitrary waveform generator
Table of Contents

Advertisement

Adjustment Procedures
5-12
2. Set up the AWG400-Series, and load the waveform file:
a. Reset the AWG400-Series:
Push UTILITY (front-panel)!System (bottom)!Factory Reset
(side)!OK (side).
b. Select the channel:
Push the desired channel select button CH1 (left side of the front panel)
for the channel that you want to verify.
c. Load the S130X2.WFM file for CH1.
Refer to Loading Files on page 4–10 for file loading procedures.
d. Select clock frequency:
Push SETUP (front-panel)!Horizontal (bottom)!Clock (side).
Push 2, 0, 0, and M (SHIFT+7) keys in this order, or turn the gener-
al purpose knob to set the internal clock frequency to 200 MHz.
e. Set output level and mode:
Push Vertical (bottom)! Amplitude (side).
Push 0, ., and 5 keys in this order or turn the general purpose knob to set
the amplitude to 0.5 V.
Push Output (side) ! Direct.
3. Push the RUN and CH1 OUTPUT front panel buttons.
4. Verify the glitch on the sine wave that is displayed on the oscilloscope
screen.
The next adjustment is performed when there is much glitch.
Adjust
NOTE. For the AWG420/AWG430, turn off the power switch, and remove the
CH2 A70/A72 output board (see Figure 5–6).
5. The connections and settings are same as for step 1 and step 2.
6. Push the RUN and CH1 OUTPUT front-panel buttons.
7. Adjust R0417 on the A50 sequence board to minimize the glitch on the sine
wave.
AWG410, AWG420 & AWG430 Service Manual