Troubleshooting And Repair; Error Messages - Rice Lake SURVIVOR IQ L Series Installation Manual

Lcd remote displays
Table of Contents

Advertisement

5.0

Troubleshooting and Repair

5.1

Error Messages

Table 5-1 lists error messages shown by the L-Series remote displays. The
encountered during the power-on self-test sequence; input error messages are shown only if DIP switch DSW3-2
is set on.
Message
Meaning
CL R F
Clock RAM failure
CLO ST
Clock stopped
CPU
Controller error
ER FT
Error in serial string format
ER TAR
Error in tare value
FAIL E
EEPROM failed self-test
FAIL R
RAM failed self-test
FAIL S
Loop-back self-test failed
FE DIG
Too few digits in serial string
IIC PR
I 2 C bus error
INV GN
Invalid gross/net format
INV S
Invalid DIP switch setting
LCD ER
LCD display error
NO BAU
Unknown baud rate
NO CLO
Clock error
NO CR
No <C/R> or <ETX> character No termination character received in serial string.
NO LBD
Unknown baud rate
NO RDR
Error reading serial format
NO SER
No serial communications
PASS
Self-test sequence passed OK No errors on power-on self-test for RAM and EEPROM functions.
REV xxx
Software revision number
T COLD
Too cold!
UN STR
Unknown serial format
UN LN2
Unknown serial format
Clock chip malfunctioning or not installed.
Clock chip malfunctioning or not installed.
Hardware error.
Error in serial string format received (more than 100 characters received without
a termination character).
Positive, non-NTEP tare value received. Check scale platform.
EEPROM portion of power-on self-test failed.
RAM portion of power-on self-test failed.
Serial port error (shown only if DSW3-7 is set on).
Error in serial string format received.
EEPROM or temperature probe error. Set DSW3-3 off if temperature probe not
installed.
Error in serial string format received. Remote display must be powered off, then
on again to continue.
Check switch settings at DSW4 or enable learn mode (set DSW2-8 on). Remote
display must be powered off, then on again to continue.
Error in LCD display configuration. Check DIP switches and board connections.
Unable to determine indicator baud rate.
Clock chip malfunctioning or not installed.
Unable to determine indicator baud rate (learn mode only).
Error in serial string format received.
Check serial connection from indicator to remote display (J10). Ensure indicator
is configured for continous (streaming) output.
Informational message at power-on, shows software revision number (xxx).
Sensed temperature is less than –99° (C or F).
Unable to determine indicator serial string format.
Unable to determine indicator serial string format (learn mode only).
Table 5-1. Error Messages
FAIL x
messages indicate errors
Cause / Correction
Troubleshooting and Repair
11