Disinfection Cycle Chart

Legend
Factory Set:
Hard Coded:
User Programmable:
Phase
Option
Start-up
Leak
Flush
Wash Soak
These settings can be changed by accessing the Diagnostics Menu.
These setting cannot be changed.
These setting can be changed by the user without accessing the
Diagnostics Menu.
Description
Inflate scope with air
Test for major air leak in the scope
Detergent injection
Endoscope channel flush with detergent
and water
Air purge through chamber line
Endoscope channel air purge
Clear air lines
Detergent injection
Endoscope channel flush with
detergent and water
Basin fill minimum (level sensor ignored)
Basin fill balance (sensor monitored)
Top off
Scope soaks in the basin. Basin fluid circulates
through channels
Drains the basin while flushing the scope
channels
Drains the rest of the fluid
Air purge through chamber line
Air purge through re-circulation lines
Endoscope channel air purge
Clear air lines
Min.
Default
Typical
Limit
20 sec
20 sec
20 sec
20 sec
3 sec
3 sec
0 sec
30 sec
30 sec
0 sec
5 sec
5 SEC
30 sec
30 sec
0 sec
5 sec
5 sec
3 sec
3 sec
0 sec
30 sec
30 sec
90 sec
90 sec
0
5 min
10 sec
30 sec
30 sec
0 sec
60 sec
60 sec
0 sec
90 sec
90 sec
0 sec
30 sec
30 sec
5 sec
5 sec
10 sec
10 sec
30 sec
30 sec
0 sec
5 sec
5 sec
Max.
State
Setting
Limit
#
Hard Coded
3
Hard Coded
4
59 sec
User
16
programmable
99m:59s
User
17
programmable
Hard Coded
92
999 sec
Factory Set
94
Hard Coded
95
User
59 sec
7
programmable
Hard Coded
118
999 sec
Factory Set
119
Hard Coded
120
999 sec
Factory Set
121
99m:59s
User
122
programmable
999 sec
Factory Set
96
Hard Coded
97
Hard Coded
92
Hard Coded
93
999 sec
Factory Set
94
Hard Coded
95
Function
#
Setup 5
Detergent
Inject
Setup 5
Flush
Diag. 63
Setup 5
Detergent
Inject
Diag 62
Diag. 67
Setup 5
Soak
Diag 61
Diag 63