Bante Instruments PHscan40BNC Instruction Manual

Bante Instruments PHscan40BNC Instruction Manual

Pocket ph tester

Advertisement

Quick Links

Hazardous Substance Statement

Bante is committed to the reduction and eventual elimination of all hazardous
substances in both the manufacturing process and finished products we supply.
We have an active manufacturing and procurement program to minimize and
eliminate the use of harmful heavy metals such as cadmium, lead, mercury and
the like.
New technologies and design parameters are also promoting these efforts and
we expect to have little or no such materials in our product in the coming years.
We welcome our customers suggestions on how to speed up these efforts.

Warranty

The warranty period for PHscan series tester is 12 months from the date of
shipment. Above warranty does not cover pH electrode module and calibration
solutions. For more information, please contact nearest authorized distributor.
BANTE INSTRUMENTS LIMITED
E-mail: banteinstruments@yahoo.com
Phone: 0086-21-6404-1598
Website: www.bante-china.com
The information in this document is subject to change without notice.
Copyright © Bante Instruments Limited, 2013. All rights reserved.
PHscan40BNC Pocket pH Tester
Instruction Manual
BANTE INSTRUMENTS LIMITED
Technology to create the future...

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PHscan40BNC and is the answer not in the manual?

Questions and answers

Summary of Contents for Bante Instruments PHscan40BNC

  • Page 1 For more information, please contact nearest authorized distributor. BANTE INSTRUMENTS LIMITED E-mail: banteinstruments@yahoo.com Phone: 0086-21-6404-1598 Website: www.bante-china.com BANTE INSTRUMENTS LIMITED The information in this document is subject to change without notice. Technology to create the future... Copyright © Bante Instruments Limited, 2013. All rights reserved.
  • Page 2 PHscan40 PHscan40 Introduction Inserting the Batteries Thank you for selecting the PHscan40 pocket pH tester. This manual provides Before using the tester, insert batteries into the battery compartment, follow a step-by-step guide to help you operate the meter, please carefully read the the procedure outlined below.
  • Page 3 PHscan40 PHscan40 Prior to Use System Menu — Remove the protective cap from the bottom of the pH electrode. The tester contains an integrated system menu that allows you to customize — If the electrode bulb dries out, soak the electrode in storage solution or tap each displayed option to meet personal preference.
  • Page 4 PHscan40 PHscan40 1. Press and hold the CAL key for 3 seconds to enter system menu, the tester Press ENTER key to confirm the tester goes to temperature setting mode, shows default pH buffer option. the display shows “ C/CAL”. Press CAL key again to select the available standard.
  • Page 5 PHscan40 PHscan40 Temperature Compensation 8. Press ENTER key to confirm, the tester shows “YES/OFF” indicating that the auto-off function is enabled. In order to get accurate results, you need to enable the manual temperature compensation function before pH measurement or calibration. 1.
  • Page 6 PHscan40 PHscan40 Single Point Calibration — Rinse the electrode with distilled water, dip the electrode into the pH4.01 1. Make sure that you have selected 1 point calibration in the system menu. or 10.01 (pH4.01 or 9.18) standard buffer solution. 2.
  • Page 7 PHscan40 PHscan40 EXIT THE CALIBRATION: — Clogged reference junction: During the calibration, if you want to exit calibration mode. Press MEAS key, Heat a diluted KCl solution to 60 to 80 C, place the sensing part of the sensor the tester will immediately return to measurement mode. into the heated solution for about 10 minutes, allow the electrode to cool in some unheated KCl solution.
  • Page 8 Temperature ( C) Replace the batteries. 4.01 6.98 9.46 4.01 6.95 9.40 Specifications 4.01 6.92 9.33 4.00 6.90 9.28 Model PHscan40BNC 4.00 6.87 9.23 pH Range -1.00~15.00pH 4.01 6.86 9.18 Resolution 0.01pH 4.01 6.85 9.14 Accuracy ±0.01pH 4.02 6.84 9.11...