PHscan20S/F Overview Thank you for selecting the PHscan20 pocket pH tester. This manual provides a step-by-step guide to help you operate the instrument, please carefully read the following instructions according to the model you have purchased. Installing the Batteries Remove the battery compartment lid.
Page 3
PHscan20S/F pH Calibration The PHscan20 tester allows 2 points calibration. We recommend calibrating the instrument during the first use or new electrode replaced. For better accuracy, regular calibration is recommended. DO NOT reuse the buffer solutions after calibration, contaminants in solution will affect the calibration and eventually the accuracy of the measurement.
Page 4
PHscan20S/F Measurements pH Measurement Rinse the electrode with deionized water. Place the electrode into the sample solution and stir gently. Note that the end of the electrode must be completely immersed into the solution. Wait for the measurement to stabilize and record the reading.
Page 5
PHscan20S/F Appendix Preparation of pH Buffer Solutions The PHscan20 tester is packaged with pH4.01/7.00/10.01 buffer sachets required for calibration. • Open the pH7.00 buffer sachet, pour the reagent into a 250 ml volumetric flask. Fill the deionized water to the mark and mix the solution until the reagent is completely dissolved.
Page 6
2002/96/EC and may not be disposed of in domestic waste. Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and electronic equipment. Bante Instruments Inc. Office: 4715 Castlewood St., Sugar land, TX 77479, USA Tel: (+1) 346-762-7358 ...