ECscan10H Pocket Conductivity Tester Introduction Thank you for selecting the ECscan10H pocket conductivity tester. This manual provides a step-by-step guide to help you operate the meter, please carefully read the following instructions before use. Inserting the Batteries Take out the G13A button cell batteries from the packaging. Follow the steps below to insert batteries into the battery compartment.
Page 3
Calibration ECscan10H pocket conductivity tester allows 1 point calibration. To ensure accurate measurement, we recommend that calibrating the meter regularly. The tester acceptable calibration ranges from 10 to 18mS/cm. Ensure that you use fresh conductivity standard solution during the calibration. DO NOT reuse calibration solutions as it may be contaminated and affect the calibration and accuracy of measurement.
Page 4
ECscan10H Pocket Conductivity Tester Ensure the displayed value matches your calibration standard, press MEAS key to confirm. The calibration value will automatically flash 3 times, the tester returns to measurement mode. Calibration is completed. Meas Hold EXIT THE CALIBRATION: During the calibration process, if you want to exit calibration mode, DO NOT press MEAS key in the step 5. Press or key until the display shows “---“, then press MEAS key to exit.
Page 5
ECscan10H Pocket Conductivity Tester Electrode Replacement When the tester fails to calibrate or gives fluctuating readings for calibration standards, you need to replace the electrode module. Twist the electrode collar counter clockwise, pull the old electrode module away from the tester.
Page 6
ECscan10H Pocket Conductivity Tester Specifications Model ECscan10H Range 0.1~19.99mS/cm Accuracy ± 1% F.S Calibration Points 1 point Calibration Solutions 12.88mS/cm Temperature Compensation 0~50° C Temperature Coefficient 2%/° C Cell Constant Normalization Temperature 25° C Hold Function Manual Power Off Manual or Automatic (8 minutes after last key pressed)
Page 7
Hazardous Substance Statement Bante Instruments Limited is committed to the reduction and eventual elimination of all hazardous substances in both the manufacturing process and finished products we supply. We have an active manufacturing and procurement program to minimize and eliminate the use of harmful heavy metals such as cadmium, lead, mercury and the like.