  • Page 1 PHscan30S/F Pocket pH Tester Instruction Manual Bante Instruments Inc.
  • Page 2: Installing The Batteries

    PHscan30S/F Overview Thank you for selecting the PHscan series pocket pH tester. This manual provides a step-by-step guide to help you operate the instrument, please carefully read the following instructions before use. Installing the Batteries Twist the electrode collar counter clockwise, pull the electrode away from the tester.
  • Page 3: Switching The Tester On And Off

    PHscan30S/F Switching the Tester On and Off Press and hold the Meas key for about 5 seconds to switch on the tester. • • Press and hold the Meas key to switch off the tester. • To enable the auto-power off function, refer to the Setup Menu section.
  • Page 4: Single Point Calibration

    PHscan30S/F pH Calibration The PHscan30 tester allows up to 3 points pH calibration, we recommend that you perform at least 2 point calibration for accurate measurement, the tester will automatically recognize and calibrate to following standard buffer values. USA standard buffers pH 4.01, 7.00, 10.01...
  • Page 5: Temperature Calibration

    PHscan30S/F 3 Points Calibration Make sure that you have selected 3 points calibration in the setup menu. Repeat steps 1.1 through 1.3 above. When the first calibration point is completed, the display will show "pH 4.01/CAL 2", the tester prompts you to continue with second point calibration.
  • Page 6: Electrode Maintenance

    PHscan30S/F Electrode Maintenance • Since pH electrode is susceptible to contamination, thoroughly clean with deionized water as necessary after each use. • If your samples contain the oil or grease, soak the electrode in mild detergent or electrode cleaning solution for at least 15 minutes.
  • Page 7: Specifications

    PHscan30S/F Solutions Order Code Description PHCS-USA pH 4.01/7.00/10.01 buffer solutions, 480 ml PHCS-OG Electrode cleaning solution, removes oil and grease contaminants, 480 ml PHCS-PR Electrode cleaning solution, removes protein contamination, 480 ml PHCS-ES pH electrode storage solution, 480 ml Specifications...
  • Page 8 Bante Instruments Inc. Office: 4715 Castlewood St., Sugar land, TX 77479, USA Tel: (+1) 346-762-7358 E-mail: banteinstruments@yahoo.com Factory: A10, No.2066, Laifang Rd., Shanghai 201615, China Tel: (+86) 21-6404-1598 E-mail: banteinstrument@hotmail.com  www.bante-china.com...