Reach-in co2 incubator 29 cu ft capacity (64 pages)
Summary of Contents for Forma Scientific 3546
Page 1
P.O. Box 649 Marietta, Ohio 45750 U.S.A. Telephone: (740) 373-4763 Telefax: (740) 373-4189 ________________________________________ Models 3546 and 3548 Water Jacketed Incubator Operating Manual Manual No. 7053546 Rev. 0 Read this Instruction Manual Failure to read, understand and follow the instructions in this manual may result in...
Page 2
In no event shall Forma Scientific, Inc. be held liable for any damages, direct or incidental, arising out of or related to the use of this manual. When contacting the factory regarding this unit, have the model, serial number and date of purchase available.
Forma Scientific, Inc.___________________________________________________Contents Table of Contents Section 1 - Receiving 1.1 Preliminary Inspection ................. 1-1 1.2 Visible Loss or Damage ............... 1-1 1.3 Responsibility for Shipping Damage ............ 1-1 Section 2 - Unpacking List 2.1 Unpacking List..................2-1 Section 3 - Installation and Start-Up 3.1 Location ....................
Page 4
Forma Scientific, Inc.___________________________________________________Contents 3.15 Connecting the gas supply when not using the Gas Guard ....3-13 3.16 Zeroing the CO 2 Controller..............3-14 3.17 Setting the CO 2 Content ..............3-15 Section 4 - Operation 4.1 Operation ..................... 4-1 4.2 Overview of Humidification and CO 2 ..........4-1 4.3 Control Panel ..................
Page 5
Forma Scientific, Inc.___________________________________________________Contents Section 5 - Routine Maintenance 5.1 Installing The Decontamination Kit ............. 5-1 5.2 Disinfecting the Incubator Interior ............5-3 5.3 Cleaning the Cabinet Exterior .............. 5-4 5.4 Draining the Water Jacket ..............5-4 5.5 Changing the CO 2 Filter ..............5-5 5.6 CO 2 Test Instruments ................
Page 6
Forma Scientific, Inc.___________________________________________________Contents Section 8 - Parts List Section 9 - Schematics Section 10 - Warranty Information...
Inc. ends when the merchandise is loaded onto the carrier's vehicle. On F.O.B. Destination shipments, Forma Scientific's and the carrier's responsibility ends when your Receiving Department personnel sign a free and clear delivery receipt. Whenever possible, Forma Scientific, Inc. will assist in settling claims for loss or in- transit damage.
Forma Scientific, Inc.________________________________________________Unpacking Section 2 - Unpacking List 2.1 Unpacking List Remove packing box from incubator. If the unit is to be moved by fork lift, leave the incubator on the skid until it has been moved to its designated location. A plastic bag containing the following accessories is packed inside the incubator: Qty.
Before installing the duct sheets and the shelves, remove the clear plastic film from the shelf brackets and duct sheets. Forma Scientific, Inc. recommends disinfecting all interior surfaces (including both door gaskets). Rinse the surfaces with sterile distilled water (50K Ohm to 1 Meg Ohm).
Forma Scientific, Inc.__________________________________________Installation/Start-up 3.3 Installing the #6 Neoprene Stopper (Access Port) Open the incubator outer door and inner glass door. Locate the opening in the top left corner of the interior chamber. Place the beveled end of the stopper in the opening.
Page 11
Forma Scientific, Inc.__________________________________________Installation/Start-up Installing the Shelf Bracketrs and Duct Sheets Figure 3-1...
Forma Scientific, Inc.__________________________________________Installation/Start-up 3.5 Installing the Duct Sheets (Refer to Figure 3-1) Note: The left duct sheet has large notches in both top and bottom edges. The right duct sheet has a single notch in the top edge only. Carefully put the right duct sheet into the incubator chamber with the flanges toward the wall.
Specification section or Electrical Schematics in the back of this manual for specific power requirements. Note: Forma Scientific, Inc. recommends that the incubator be connected to a separate electrical circuit. 3.9 Preparing the Incubator for Filling 450 ml of rust inhibitor was placed in the water jacket before the incubator was shipped.
Page 14
Forma Scientific, Inc.__________________________________________Installation/Start-up Turn the incubator power switch ON. The ADD WATER warning light and audible alarm will come on. Pressing the SET/SILENCE button on the alarm panel will silence the audible alarm Remove the adjusting screwdriver from the front of the control panel by pulling out the black knob beneath the Sample Port.
Forma Scientific, Inc.__________________________________________Installation/Start-up Press the CO 2 SET/SILENCE button and rotate the CO 2 set screw until the display reads 00.0. (Refer to Figure 3-6 for locations of the SET/SILENCE button and set screw.) 3.10 Filling the Water Jacket There are two methods of filling the water-jacket.
Forma Scientific, Inc.__________________________________________Installation/Start-up Add one additional quart (one liter) of distilled water. The incubator is now properly filled. Remove the tubing from the fill fitting and replace the plastic protective cap. Note: Water seepage may occur from vent hole when chamber temperature increases.
Forma Scientific, Inc.__________________________________________Installation/Start-up Turn the water off immediately when the ADD WATER alarm light goes out. Using the funnel, add an additional quart (one liter) of distilled water. Remove vinyl tubing and reinstall the plastic protective cap. 3.11 Filling the Humidity Reservoir or Pan Do not use plastic pans for humidification as they will have an unpredictable effect on humidity and O 2 /CO 2 levels in the incubator.
Forma Scientific, Inc.__________________________________________Installation/Start-up When installing the humidity pan, exercise care to avoid tearing the inner door gasket. 3.12 Setting the Chamber Temperature (Refer to Figure 3-4) Before the initial temperature setting is made, push in on the PUSH TO SET button on the Alarm Monitor Module.
Page 19
Forma Scientific, Inc.__________________________________________Installation/Start-up Temperature and Alarm Monitor Module Figure 3-4 3-11...
30 PSIG to monitor actual input pressure to the incubator injection system. A suitable two-stage pressure regulator is available from Forma Scientific, Inc., Stock #965010. Supply pressure lower than 13.5 PSIG will cause improper operation of the gas guard and may cause nuisance alarms.
Forma Scientific, Inc.__________________________________________Installation/Start-up 3.15 Connecting the Gas Supply when not using the Gas Guard (Refer to Figure 3-6) If the gas guard is not to be used, the gas supply must be connected to the CO 2 inlet. To provide for the most economical use of CO 2 , a main supply of liquid CO 2 is recommended.
Forma Scientific, Inc.__________________________________________Installation/Start-up 3.16 Zeroing the CO 2 Controller (See Figure 3-7) This adjustment is made using the CO 2 gas content of ambient air (0.03%), the most accurate standard available. Never use a FYRITE or other analyzer for this adjustment. The adjustment must be made on initial start-up, and it must also be made if a change in the humidification of the incubator is required.
Forma Scientific, Inc.__________________________________________Installation/Start-up Check the CO 2 at the Desired Setpoint Allow the incubator to reach setpoint and control (inject light will cycle) for a minimum of 30 minutes. Check the CO 2 level with a FYRITE until two consecutive readings agree. If the FYRITE and display are not within plus or minus 1.0%, consult the factory.
Forma Scientific, Inc.__________________________________________________Operation Section 4 - Operation 4.1 Operation The water jacket is filled with 11.5 gallons (43.5 liters) of water through the fill fitting located on the front of the unit. The water is then warmed by the chamber heater, providing very stable heating of the incubator chamber.
Page 25
Forma Scientific, Inc.__________________________________________________Operation When monitoring the effects of CO 2 , absolute humidity must be held constant so any change in thermal conductivity is caused only by a change in the CO 2 concentration. Under the worst circumstances, a change in absolute humidity could cause such a significant change in thermal conductivity that the controller could shift the CO 2 content by as much as 4%.
Forma Scientific, Inc.__________________________________________________Operation 4.3 Control Panel (Figure 4-2) Fill Fitting and Vent Hole The fill fitting is located on the center front of the unit, directly above the door. This fitting is to facilitate filling of the water jacket without having to move the unit.
Forma Scientific, Inc.__________________________________________________Operation Variable/37°C Switch, Temp Control and Heater Pilot Light When the Variable/37°C switch is set to the 37°C (up) position, chamber temperature will automatically be maintained at +37°C. If necessary, the 37°C control can be calibrated using the calibration screw located at the lower left side of the temperature control dial.
Forma Scientific, Inc.__________________________________________________Operation Control Module Figure 4-3 Audible CO 2 Alarm and Pilot Light The audible CO 2 alarm and pilot light are activated when the percent CO 2 deviates from setpoint by plus or minus 1% (nominal) for longer than approximately four minutes.
Forma Scientific, Inc.__________________________________________________Operation Figure 4-4 Alarm Disable Switch on the CO 2 Control Module CO 2 Inject Light The CO 2 inject light is activated whenever there is a demand for CO 2 to meet setpoint requirements. Since the CO 2 inject light is independent of the CO 2 alarm, it will continue to signal a need for CO 2 when the CO 2 alarm is set to either the defeat or silence position.
Forma Scientific, Inc.__________________________________________________Operation The CO 2 zero adjustment is used for zeroing the CO 2 controller to specific control conditions. It is the only user calibration adjustment on the CO 2 module. All internal adjustments are for qualified service personnel only.
Page 31
REMOTE ALARM RELAY WIRING DIAGRAM When the CONTROL switch on the gas guard is turned off the indicator light will be extinguished and the audible alarm will be silenced. The built-in alarm relay on the Model 3546, however, remains active. Figure 4-7...
Recorder Output Jack 4.8 (Optional) Gas Guard, Stock #190410 (Refer to Figure 4-9) Forma Scientific's stock # 190410 built-in Gas Guard is a monitoring and switching device for incubator gas supplies. The unit monitors a two cylinder gas supply. When the active supply tank is depleted,...
Forma Scientific, Inc.__________________________________________________Operation Control Switch The CONTROL switch controls power to the gas guard control circuit board. The yellow LEDs on the front panel designate which tank is active and indicate the unit is receiving power. Silence Switch The SILENCE switch silences the alarm and selects the alternate tank.
Forma Scientific, Inc.________________________________________________Maintenance 5.2 Disinfecting the Incubator Interior Use an appropriate disinfectant. All articles and surfaces to be disinfected must be thoroughly cleaned and rough dried. Danger: Alcohol, even a 70% solution, is volatile and flammable. Use it only in a well ventilated area that is free from open flame.
Forma Scientific, Inc.________________________________________________Maintenance After complete decontamination, it is recommended that the incubator be run and tested before placing any valuable contents inside the incubator chamber. 5.3 Cleaning the Cabinet Exterior The incubator exterior may be cleaned with soap and water and a general use laboratory disinfectant.
Forma Scientific, Inc.________________________________________________Maintenance 5.5 Changing the CO 2 Filter (Qualified Service Personnel Only) Note: Under conditions of normal use, and dependent upon the purity of gas being used, the CO 2 filter should be serviceable for about five years. If the CO 2 filter becomes clogged, replace it with Forma stock #770001.
Forma Scientific, Inc._____________________________________________Service Section 6 - Service 6.1 General Troubleshooting Symptom Possible Cause • Problem 1: Pilot lights not on. Readouts Unit unplugged, or main power switch not on. • are dark. CO 2 power switch is on. Circuit breaker tripped.
Page 41
Forma Scientific, Inc._____________________________________________Service Symptom Possible Cause Problem 7: Digital CO 2 display and • Alarm is disabled. • FYRITE read more than 1% different from Defective alarm circuit. • setpoint. CO 2 alarm not on. Defective CO 2 sampler or control.
Page 42
Forma Scientific, Inc._____________________________________________Service Symptom Possible Cause • Problem 12: CO 2 setpoint cannot be Defective CO 2 control. changed. • Problem 13: Unit cannot be zeroed. Defective CO 2 potentiometer. • Defective zero control. • Defective CO 2 Sensor. •...
Forma Scientific, Inc._____________________________________________Service 6.2 CO 2 Control Module Calibration These calibration procedures will be necessary only under the following conditions. DO NOT calibrate the incubator for any other reasons! ♦ After the CO 2 controller is replaced. ♦ After the CO 2 sensor is replaced.
Forma Scientific, Inc._____________________________________________Service CO 2 Control Module showing the Extender Card and Span Adjustment Pot Location Figure 6-1 a. Stabilizing the Incubator at operating temperature and humidity with no CO 2 in the chamber. Turn OFF the CO 2 supply at the source.
Forma Scientific, Inc._____________________________________________Service Side View of the CO 2 Control Module Figure 6-2 c. Adjusting the Span Pot (Refer to Figure 6-2) Turn ON the CO 2 at the supply. Turn the CO 2 setpoint to 10%. Allow the CO 2 to stabilize at 10% on the readout and control (inject light will cycle) for a minimum of 15 minutes.
Forma Scientific, Inc._____________________________________________Service d. Re-checking the Zero Adjustment Turn OFF the CO 2 at the supply. Open both doors wide for 45 seconds. Close the doors, and allow a minimum of 15 minutes for the incubator to stabilize and assure a zero CO 2 condition in the chamber.
Forma Scientific, Inc._____________________________________________Service 6.3 37°C Control Calibration If the digital display indicates a temperature other than 37°C when the chamber temperature has stabilized and the temperature select switch is set to 37°C, it may be necessary to calibrate the adjustment.
Forma Scientific, Inc._____________________________________________Service After the sensor has been replaced, calibrate the CO 2 controller according to the instructions in Section 6.2. 6.5 Replacing the Triac (Refer to Wiring Diagram 140179-71-0-D) Turn the incubator off and disconnect it from the power source.
Forma Scientific, Inc._____________________________________________Service 6.7 Replacing the Temperature Control (Refer to Wiring Diagram, 140179-71-0-D) Turn the incubator off and disconnect it from the power source. Remove the screws securing the right side access panel. Remove the panel and set aside. Remove the temperature control knob by loosening the two Allen head screws recessed in the knob.
Forma Scientific, Inc._____________________________________________Service 6.9 Replacing the Circuit Breaker Turn the incubator off and disconnect it from the power source. Remove the screws securing the right side access panel. Remove the panel and set aside. Remove the two nuts securing the circuit breaker mounting plate to the control panel.
Forma Scientific, Inc._____________________________________________Service Cut the wires to the defective thermistor, making note of the connections. Attach the new thermistor using the electrical in-line connectors. Install the new thermistor by reversing the above procedure. Take care not to damage the probe tip and make sure that the probe is fully extended in the sheath.
Page 52
Forma Scientific, Inc._____________________________________________Service Remove the four wing nuts that secure the blower mounting plate to the incubator ceiling. Pull the blower motor assembly down into the chamber area. Remove and save the upper blower wheel/blower blade. Cut the tie wrap(s) securing the blower wires to the blower shield plate.
Forma Scientific, Inc._____________________________________________Service 6.14 Fuse Chart Fuse Manufacturer’s Part No. Amp Rating Rupture Speed Control Buss MDL-3/10 0.3 Amp Time Lag 230VAC Buss MDL-3 3.0 Amp Time Lag (3548 only) 6-14...
Forma Scientific, Inc.______________________________Specifications/Accessories Section 7 - Features, Specifications and Accessories Models 3546/3548 7.1 Construction • Fiberglass insulation combines with the water jacket to form triple-wall protection of the chamber. • Smooth, seam-welded stainless steel interior provides a crevice-free chamber, reducing contamination.
Forma Scientific, Inc.______________________________Specifications/Accessories 7.3 Temperature Control • Chamber temperature is constantly displayed on easy-to-read LCD. • Independent, electronic, dual temperature controller maintains chamber temperature at a fixed +37 o C or adjustable from +5 o C above ambient to +60 o C.
Need help?
Do you have a question about the 3546 and is the answer not in the manual?
Questions and answers