LIFEPAK 1000
®
DEFIBRILLATOR
OPERATING INSTRUCTIONS
To Purchase, Visit
Avobus.com
or call
1-800-674-3655

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the LIFEPAK 1000 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Physio Control LIFEPAK 1000

  • Page 1 LIFEPAK 1000 ® DEFIBRILLATOR OPERATING INSTRUCTIONS To Purchase, Visit Avobus.com or call 1-800-674-3655...
  • Page 3 LIFEPAK 1000 ® DEFIBRILLATOR OPERATING INSTRUCTIONS...
  • Page 4 Important This instrument is to be used by authorized personnel only. Device Registration Please register your device at www.physio-control.com. This will ensure that you are notified of any product updates. Responsibility for Information It is the responsibility of our customers to ensure that the appropriate person(s) within their organization have access to this information, including cautions and warnings provided throughout this manual.
  • Page 5: Table Of Contents

    3 How to Use the LIFEPAK 1000 Defibrillator Modes of Operation ....................3-2 Defibrillation Warnings and Cautions ..............3-2 Defibrillation in AED Mode ..................3-3 Basic Steps for Using the LIFEPAK 1000 Defibrillator ..........3-3 Voice Prompts and Messages in AED Mode ............3-4 LIFEPAK 1000 Defibrillator Operating Instructions ©2006-2016 Physio-Control, Inc.
  • Page 6 4 Data Management Managing Defibrillator Data ..................4-2 Overview of Data Storage..................4-2 Data Stored by the LIFEPAK 1000 Defibrillator............ 4-2 Overview of Connections for Transmitting Reports ..........4-3 5 Caring for the LIFEPAK 1000 Defibrillator Maintenance and Testing Schedule ................5-2 Self-Test Performance ....................
  • Page 7 This section provides information about defibrillation and an overview of the LIFEPAK ® 1000 defibrillator. About Defibrillation page vi Indications for Use Operator Considerations About the LIFEPAK 1000 Defibrillator Text Conventions viii LIFEPAK 1000 Defibrillator Operating Instructions To Purchase, Visit Avobus.com or call 1-800-674-3655...
  • Page 8: Preface

    25 kg (55 lbs) with Infant/Child Reduced Energy Defibrillation Electrodes. ECG Monitoring ECG monitoring is for use on conscious and unconscious patients of all ages for the purpose of ECG rhythm recognition and heart rate monitoring. Contraindications None known. LIFEPAK 1000 Defibrillator Operating Instructions...
  • Page 9: Operator Considerations

    ABOUT THE LIFEPAK 1000 DEFIBRILLATOR The LIFEPAK 1000 defibrillator is a semiautomatic model that can be operated in either of three modes: AED mode, Manual mode, and ECG mode. The defibrillator uses the Physio-Control patented Shock Advisory System™ to analyze the patient's electrocardiographic (ECG) rhythm and prompts you when it detects a shockable rhythm and when it does not detect a shockable rhythm.
  • Page 10: Text Conventions

    Readiness Display The LIFEPAK 1000 defibrillator includes a readiness display. The OK symbol appears in the display if the daily self-test is completed successfully. A battery symbol that approximates the remaining state of charge is also visible.
  • Page 11: Safety

    SAFETY This section provides important information to help you operate the LIFEPAK 1000 defibrillator. Familiarize yourself with all of these terms, warnings, and symbols. Terms page 1-2 General Warnings and Cautions Symbols LIFEPAK 1000 Defibrillator Operating Instructions ©2018 Physio-Control, Inc.
  • Page 12: Terms

    Safety TERMS The following terms are used either in this manual or on the LIFEPAK 1000 defibrillator. Danger: Immediate hazards that will result in serious personal injury or death. Warning: Hazards or unsafe practices that could result in serious personal injury or death.
  • Page 13: Symbols

    If the device has been abused, remove it from use and contact authorized service personnel. SYMBOLS The following symbols may be found in this manual or on various configurations of the LIFEPAK 1000 defibrillator and its accessories. Defibrillation-proof type BF applied part Operating instructions Caution Warning.
  • Page 14 Canadian Standards Association certification mark for Canada and the United States Underwriters Laboratories recognized component mark for Canada and the United States Not made with natural rubber latex Cable Connector !USA For USA audiences only Manufacturer Date of manufacture shown: YYYY-MM-DD LIFEPAK 1000 Defibrillator Operating Instructions...
  • Page 15 FIRE OR BURN HAZARD - Do not deform or damage EXPLOSION HAZARD - Do not dispose of in fire or heat above 100°C (212°F) FIRE OR BURN HAZARD - Do not open or dismantle LIFEPAK 1000 Defibrillator Operating Instructions ©2018 Physio-Control, Inc.
  • Page 16 Non-rechargeable battery. Do not charge. Lithium manganese dioxide battery Li MnO Rechargeable battery Battery for use with the LIFEPAK 1000 defibrillator 1000 Battery charger for use with Lithium-ion battery AC to DC power adapter Do not dispose of this product in the unsorted municipal waste stream.
  • Page 17: Controls And Indicators

    CONTROLS AND INDICATORS This section provides a description of each of the LIFEPAK 1000 defibrillator primary controls and indications. Controls and Indicators page 2-2 Battery Indicators Battery Charger Indicators LIFEPAK 1000 Defibrillator Operating Instructions ©2018 Physio-Control, Inc.
  • Page 18 Controls and Indicators CONTROLS AND INDICATORS This section introduces you to the controls and indicators on the LIFEPAK 1000 defibrillator. (Left side) (Left side) Controls and Indicator Figure 2-1 Controls and Indicators Table 2-1 Feature Description Readiness display The readiness display alerts you to the defibrillator’s readiness status.
  • Page 19 Cable receptacle Allows direct connection to therapy electrodes (black), ECG cable (green), Infant/Child electrodes (pink), and QUIK-COMBO therapy electrodes (gray). LIFEPAK 1000 Defibrillator Operating Instructions To Purchase, Visit Avobus.com or call 1-800-674-3655 ©2018 Physio-Control, Inc.
  • Page 20 Lead II ECG cable. The presence of an ECG does not ensure that the patient has a pulse. Softkey labels. These labels define the function that can be activated by pressing the softkey. ANALYZE and DISARM are function examples. LIFEPAK 1000 Defibrillator Operating Instructions...
  • Page 21: Battery Indicators

    Controls and Indicators BATTERY INDICATORS The LIFEPAK 1000 defibrillator can be powered by two types of batteries: • A nonrechargeable Lithium manganese dioxide battery • A rechargeable Lithium-ion battery Battery Charge Level Indicators Both battery types have a fuel gauge that indicates the approximate charge level of the battery when it is not installed in a defibrillator.
  • Page 22: Battery Charger Indicators

    BATTERY CHARGER INDICATORS The LIFEPAK 1000 Defibrillator Battery Charger is intended for use with LIFEPAK 1000 defibrillator Lithium-ion (Li-ion) rechargeable batteries. No other batteries are compatible with this charger. For complete battery charger information, refer to the LIFEPAK 1000 Defibrillator Battery Charger Instructions for Use provided with the charger.
  • Page 23: How To Use The Lifepak 1000 Defibrillator

    HOW TO USE THE LIFEPAK 1000 DEFIBRILLATOR This section provides an overview of information and instructions for using the LIFEPAK 1000 defibrillator. Modes of Operation page 3-2 Defibrillation in AED Mode Defibrillation in Manual Mode Troubleshooting Tips for Defibrillation ECG Monitoring (ECG Mode) LIFEPAK 1000 Defibrillator Operating Instructions ©2018 Physio-Control, Inc.
  • Page 24: Modes Of Operation

    How to Use the LIFEPAK 1000 Defibrillator MODES OF OPERATION You can use the LIFEPAK 1000 defibrillator for: • Automated external defibrillation (AED mode) • Manual defibrillation therapy (Manual mode) (Requires ECG display option) • ECG monitoring (ECG mode) (Requires ECG display option)
  • Page 25: Defibrillation In Aed Mode

    The LIFEPAK 1000 defibrillator uses the Physio-Control patented Shock Advisory System to evaluate the patient’s heart rhythm. The LIFEPAK 1000 defibrillator has an optional feature that displays the ECG waveform and Heart Rate Indicator in AED mode. The operation in AED mode remains the same whether or not the defibrillator displays the ECG waveform.
  • Page 26: Voice Prompts And Messages In Aed Mode

    How to Use the LIFEPAK 1000 Defibrillator 6 Connect the electrodes to the defibrillator (if they are not already connected). 7 Follow the screen messages and voice prompts provided by the defibrillator. Voice Prompts and Messages in AED Mode The following descriptions of voice prompts and messages are based on the default settings for AED mode.
  • Page 27: Special Situations For Electrode Placement

    How to Use the LIFEPAK 1000 Defibrillator Special Situations for Electrode Placement When placing electrodes on the patient, be aware of special situations: Obese Patients or Patients with Large Breasts Apply the electrodes to a flat area on the chest, if possible. If skin folds or breast tissue prevent good adhesion, spread skin folds apart to create a flat surface.
  • Page 28: Defibrillation In Manual Mode

    How to Use the LIFEPAK 1000 Defibrillator DEFIBRILLATION IN MANUAL MODE The LIFEPAK 1000 defibrillator provides a Manual mode to override the AED features of the defibrillator. Manual mode provides operator-initiated analysis, charge, shock, and disarm functions. This mode is useful in a tiered response system when a provider trained in manual defibrillation and authorized to place the defibrillator in Manual mode takes over the scene from a BLS-AED trained provider.
  • Page 29: Troubleshooting Tips For Defibrillation

    How to Use the LIFEPAK 1000 Defibrillator TROUBLESHOOTING TIPS FOR DEFIBRILLATION This section explains problem conditions that you may encounter while using the defibrillator. Troubleshooting Tips for Defibrillation Table 3-1 Observation Possible Cause What To Do Screen blank and ON Screen not functioning •...
  • Page 30 How to Use the LIFEPAK 1000 Defibrillator Troubleshooting Tips for Defibrillation (Continued) Table 3-1 Observation Possible Cause What To Do MOTION DETECTED Patient movement because of • Move patient to stable location, if and STOP MOTION location. possible. voice prompts are Patient movement because •...
  • Page 31: Ecg Monitoring (Ecg Mode)

    Do not attempt to connect a 3-wire ECG cable to a QUIK-COMBO therapy cable or any other AED. The ECG cable is functional only with the LIFEPAK 1000 defibrillator. The LIFEPAK 1000 defibrillator provides nondiagnostic ECG display of the patient’s heart rhythm when the ECG cable is connected and the electrodes are applied.
  • Page 32: Troubleshooting Tips For Ecg Monitoring Mode

    How to Use the LIFEPAK 1000 Defibrillator Troubleshooting Tips for ECG Monitoring Mode If problems occur while monitoring the ECG, check this list of observations for troubleshooting assistance. Troubleshooting Tips for ECG Monitoring Mode Table 3-2 Observation Possible Cause What to Do...
  • Page 33 How to Use the LIFEPAK 1000 Defibrillator Troubleshooting Tips for ECG Monitoring Mode (Continued) Table 3-2 Observation Possible Cause What to Do Poor ECG signal quality. Poor electrode-to-skin • Reposition cable and/or lead contact. wires to prevent electrodes from pulling away from patient.
  • Page 34 To Purchase, Visit Avobus.com or call 1-800-674-3655...
  • Page 35: Data Management

    DATA MANAGEMENT This section introduces data management for the LIFEPAK 1000 defibrillator. Managing Defibrillator Data page 4-2 LIFEPAK 1000 Defibrillator Operating Instructions ©2018 Physio-Control, Inc.
  • Page 36: Managing Defibrillator Data

    The LIFEPAK 1000 defibrillator does not delete patient data after you transfer the data to a PC. The defibrillator deletes previous patient data only when it is connected to a new patient or a simulator.
  • Page 37: Overview Of Connections For Transmitting Reports

    Data Management Test and Service Data The LIFEPAK 1000 defibrillator stores a test log consisting of the most recent auto-tests, power cycles, and battery replacements. The test log lists the test results and any errors detected. The test log data is available only to authorized service personnel or to responders who are using the appropriate LIFENET system product.
  • Page 38 This includes initiating data download, selecting reports to be transmitted, and monitoring transmission progress. More information about configuring your LIFENET system product and instructions for transmitting device data are provided in the users guide and reference cards that accompany the LIFENET system product. LIFEPAK 1000 Defibrillator Operating Instructions...
  • Page 39: Caring For The Lifepak 1000 Defibrillator

    CARING FOR THE LIFEPAK 1000 DEFIBRILLATOR This section explains how to help keep your LIFEPAK 1000 defibrillator in good working condition. Cared for properly, the defibrillator is built to give you many years of service. Maintenance and Testing Schedule page 5-2...
  • Page 40: Maintenance And Testing Schedule

    SELF-TEST PERFORMANCE Whenever the LIFEPAK 1000 defibrillator is turned on after it has been off for at least 60 seconds, it takes approximately 5 seconds to complete a self-test and to indicate a low or replace battery condition.
  • Page 41: Auto Tests

    Caring for the LIFEPAK 1000 Defibrillator Auto Tests The defibrillator performs automatic self-tests daily and monthly at 0300 (3:00 a.m.) if not in use. During the automatic self-test, the defibrillator turns itself on (ON/OFF LED illuminates) briefly and completes the following tasks: •...
  • Page 42: Cleaning

    • Peroxide (peracetic acid) solutions BATTERY MAINTENANCE The LIFEPAK 1000 defibrillator can be powered by either of two types of batteries: • A nonrechargeable Lithium manganese dioxide battery • A rechargeable Lithium-ion battery Follow the guidelines described in this section to help maximize battery life and performance.
  • Page 43: Lifepak 1000 Defibrillator Nonrechargeable Battery

    The Lithium-ion (Li-ion) rechargeable battery is the appropriate battery option when the LIFEPAK 1000 defibrillator is used on a frequent basis, or is used with a simulator for training. When optimally maintained, a new fully-charged battery can provide approximately ten hours of “ON”...
  • Page 44 • Recharge when the battery fuel gauge shows that the battery charge level is low. • Use only the Physio-Control battery charger designed for use with the LIFEPAK 1000 defibrillator. Do not use any other charger. Refer to the LIFEPAK 1000 Defibrillator Battery Charger Instructions for Use.
  • Page 45: Electrode Care And Storage

    Do not disassemble the defibrillator. It contains no responder-serviceable components and dangerous high voltages may be present. Contact authorized service personnel. If the LIFEPAK 1000 defibrillator requires service as indicated by testing, troubleshooting, or the 1.800.442.1142. service symbol, contact authorized service personnel. In the USA, call Outside the USA, contact your local Physio-Control representative.
  • Page 46: Product Recycling Information

    Packaging should be recycled according to national and local regulations. SUPPLIES, ACCESSORIES, AND TRAINING TOOLS Table 5-4 lists supplies, accessories, and training tools for the LIFEPAK 1000 defibrillator. To order in the USA, call 1.800.442.1142. Outside the USA, contact your local Physio-Control representative.
  • Page 47: Warranty Information

    Caring for the LIFEPAK 1000 Defibrillator Supplies, Accessories, and Training Tools (Continued) Table 5-4 Item Description 3-Wire Monitoring Cable (IEC) Quick Reference Card IrDA Adapter (attachment for a PC) CODE-STAT Data Review Software DT EXPRESS™ Data Transfer Software WARRANTY INFORMATION To obtain a detailed warranty statement, contact your local Physio-Control representative or go to www.physio-control.com.
  • Page 49 APPENDIX A SPECIFICATIONS LIFEPAK 1000 Defibrillator Operating Instructions To Purchase, Visit Avobus.com or call 1-800-674-3655 ©2018 Physio-Control, Inc.
  • Page 51 Rated energy output is the nominal delivered energy based on the energy setting and patient impedance, as defined in the following chart. Rated Energy Output Patient Impedance () * Energy setting selected LIFEPAK 1000 Defibrillator Operating Instructions ©2018 Physio-Control, Inc.
  • Page 52 (Joules) 50.3 20.1 19.7 10.7 28.2 14.6 14.5 19.8 11.7 11.7 15.5 10.0 12.9 10.4 11.1 11.1 11.7 12.2 Table values are nominal for a 200-joule shock. Note: LIFEPAK 1000 Defibrillator Operating Instructions To Purchase, Visit Avobus.com or call 1-800-674-3655...
  • Page 53 1 5 0 1 7 5 P a t i e n t Im p e d a n c e ( o h m s ) Patient Impedance () * Energy setting selected LIFEPAK 1000 Defibrillator Operating Instructions ©2018 Physio-Control, Inc.
  • Page 54 Waveshape and Measured Parameters: Patient I1 (A) I2 (A) I3 (A) I4 (A) T1 (ms) T2 (ms) Impedance () 19.4 10.2 10.1 13.2 10.1 Note: Table values are nominal for a 50-joule shock (200 ÷ 4). LIFEPAK 1000 Defibrillator Operating Instructions...
  • Page 55 Applies for a new fully charged battery or after shock ready time seconds seconds 15 full energy discharges. (AED mode) *Device behavior may vary for Arabic, Finnish, Icelandic, and Slovenian. LIFEPAK 1000 Defibrillator Operating Instructions To Purchase, Visit Avobus.com or call 1-800-674-3655 ©2018 Physio-Control, Inc.
  • Page 56 Service required when displayed Environmental All performance specifications defined assume that the device has been stored Note: (two hours minimum) at the operating temperature prior to operation. Operating 0° to 50°C (32° to 122°F) Temperature LIFEPAK 1000 Defibrillator Operating Instructions...
  • Page 57 • Event Log report—Report of time-stamped markers reflecting operator and device activity • Test Log report—Device self-test activity report Capacity Minimum 100 time-stamped event log entries Data Review CODE-STAT 6.0 (minimum) or DT EXPRESS 2.0 (minimum) software LIFEPAK 1000 Defibrillator Operating Instructions ©2018 Physio-Control, Inc.
  • Page 58 A new fully-charged battery provides device power for 6 months. Low Battery At least 30 200-joule shocks or 75 minutes of operating time Indicator remain when low battery is first indicated. LIFEPAK 1000 Defibrillator Operating Instructions To Purchase, Visit Avobus.com or call 1-800-674-3655...
  • Page 59 APPENDIX B SHOCK ADVISORY SYSTEM LIFEPAK 1000 Defibrillator Operating Instructions ©2018 Physio-Control, Inc.
  • Page 61: B Shock Advisory System

    The Physio-Control patented Shock Advisory System (SAS™) is an ECG analysis system built into the LIFEPAK 1000 defibrillators that advises the operator whether it detects a shockable or nonshockable rhythm. This system makes it possible for individuals not trained to interpret ECG rhythms to provide potentially-lifesaving therapy to victims of ventricular fibrillation or pulseless ventricular tachycardia.The Shock Advisory System contains the following features:...
  • Page 62 • Other organized rhythms including 30 or more of each of the following rhythms: atrial fibrillation, atrial flutter, 2nd degree atrioventricular block, 3rd degree atrioventricular block, idioventricular rhythms, sinus bradycardia, supraventricular tachycardia, and rhythms with premature ventricular complexes LIFEPAK 1000 Defibrillator Operating Instructions...
  • Page 63 C. Summary Shock Advisory System Performance Report The results of tests with the SAS and Physio-Control test sets in the LIFEPAK 1000 defibrillator are shown below in the context of requirements from IEC 60601-2-4 and the recommendations from the American Heart Association.
  • Page 64 The system also disarms if the patient impedance decreases suddenly or the patient impedance goes outside the acceptable range for analysis. If none of these occur, the shockable rhythm decision is not revised during charging or prior to shock. LIFEPAK 1000 Defibrillator Operating Instructions To Purchase, Visit Avobus.com...
  • Page 65 The Shock Advisory System detects patient motion independent of ECG analysis. MOTION DETECTION can be set up to be ON or OFF. For more information, see “Changing Setup Options” in the LIFEPAK 1000 Defibrillator Operating Instructions. A number of activities can create motion, including CPR, rescuer movement, patient movement, and vehicle movement.
  • Page 67: C Cprmax™ Technology

    APPENDIX C cprMAX™ TECHNOLOGY LIFEPAK 1000 Defibrillator Operating Instructions To Purchase, Visit Avobus.com or call 1-800-674-3655 ©2018 Physio-Control, Inc.
  • Page 69 AED will prompt, IF YOU WITNESSED THE ARREST, PUSH CANCEL , which provides the opportunity to end CPR early and proceed directly to delivering a shock. LIFEPAK 1000 Defibrillator Operating Instructions ©2018 Physio-Control, Inc.
  • Page 70 Choices for the STACKED SHOCKS option are ON or OFF . The default setting is OFF . When this option is set to ON , the defibrillator follows the previously traditional stacked shock protocol and delivers up to three consecutive shocks, as necessary, without interposed CPR. LIFEPAK 1000 Defibrillator Operating Instructions To Purchase, Visit Avobus.com...
  • Page 71 NO SHOCK ADVISED decision, regardless of the first analysis decision ( SHOCK ADVISED or NO SHOCK ADVISED ). • The NEVER option eliminates all PULSE CHECK prompts. LIFEPAK 1000 Defibrillator Operating Instructions ©2018 Physio-Control, Inc.
  • Page 73 APPENDIX D CHANGING SETUP OPTIONS LIFEPAK 1000 Defibrillator Operating Instructions To Purchase, Visit Avobus.com or call 1-800-674-3655 ©2018 Physio-Control, Inc.
  • Page 75: D Changing Setup Options

    To exit Setup mode, turn the defibrillator off. If you changed the setup options, the changes are saved and will appear the next time you turn the defibrillator on. (Refer to Setup menu options that follow.) LIFEPAK 1000 Defibrillator Operating Instructions ©2018 Physio-Control, Inc.
  • Page 76 Set up Manual mode defaults. SET PASSCODE Set passcode to enter Setup mode. SERVICE MODE View service options. To choose an option, highlight your choice on the screen and press SELECT . LIFEPAK 1000 Defibrillator Operating Instructions To Purchase, Visit Avobus.com or call 1-800-674-3655...
  • Page 77 , OFF . mode. AUTO ANALYZE Select analyze option. , AFTER FIRST SHOCK, OFF . MOTION Alert when motion is detected. ON , OFF . DETECTION PREVIOUS PAGE Go back to previous page. LIFEPAK 1000 Defibrillator Operating Instructions ©2018 Physio-Control, Inc.
  • Page 78 PRESHOCK CPR Set CPR interval before , 15, 30 seconds. SHOCK ADVISED decisions. , OFF . CPR PROMPT Enable extended CPR prompt: PROVIDE RESCUE BREATHS AND CHEST COMPRESSIONS PREVIOUS PAGE Go back to previous page. LIFEPAK 1000 Defibrillator Operating Instructions...
  • Page 79 , OFF . Enable Shock Advisory ANALYZE System analysis in Manual mode. (An ANALYZE softkey is provided in Manual mode.) PREVIOUS PAGE Go back to previous page. LIFEPAK 1000 Defibrillator Operating Instructions To Purchase, Visit Avobus.com or call 1-800-674-3655 ©2018 Physio-Control, Inc.
  • Page 80 Start defibrillator calibration. PIXEL TEST Test display pixels. SERVICE LOG Show service log. SERVICE DATA Show device data. DEVICE LOG Display device log. SET PASSCODE Set Service mode access passcode. SETUP MODE Go back to Setup mode. LIFEPAK 1000 Defibrillator Operating Instructions...
  • Page 81 1 Use the clockwise or counterclockwise arrows to navigate to the BACKSPACE option. 2 Navigate to the CLEAR option and press the MENU button again. The character no longer appears on the screen. LIFEPAK 1000 Defibrillator Operating Instructions ©2018 Physio-Control, Inc.
  • Page 82 To Purchase, Visit Avobus.com or call 1-800-674-3655...
  • Page 83 APPENDIX E USER’S CHECKLIST LIFEPAK 1000 Defibrillator Operating Instructions ©2018 Physio-Control, Inc.
  • Page 85: E User's Checklist

    LIFEPAK ® 1000 DEFIBRILLATOR USER’S CHECKLIST Unit Serial Number __________________________ Department/Location ________________________ Date Recommended Initials Instruction Corrective Action Check readiness display for: symbol Contact authorized service personnel. symbol None. Battery level Replace if one bar or less on battery status indicator. Check Use By date on Replace electrode packet if date electrode packet.
  • Page 87: F Defibrillation Clinical Summaries

    APPENDIX F DEFIBRILLATION CLINICAL SUMMARIES LIFEPAK 1000 Defibrillator Operating Instructions ©2018 Physio-Control, Inc.
  • Page 88 To Purchase, Visit Avobus.com or call 1-800-674-3655...
  • Page 89: Defibrillation Of Ventricular Fibrillation And Ventricular Tachycardia

    80-96% 200 J BTE 39/39 (100%) 91-100% 130 J BTE 39/47 (83%) 69-92% S.L. Higgins et al., “A comparison of biphasic and monophasic shocks for external defibrillation,” Prehospital Emergency Care, 2000, 4(4):305-13. LIFEPAK 1000 Defibrillator Operating Instructions ©2018 Physio-Control, Inc.
  • Page 90 200 J monophasic shocks, 200 J biphasic shocks will in some cases enable earlier termination of VF. Therefore, we conclude that biphasic shocks for VF delivered at conventional energy levels have the potential to improve outcome in resuscitation of patients with cardiac arrest. LIFEPAK 1000 Defibrillator Operating Instructions...
  • Page 91: Monophasic Vs. Biphasic Waveforms: Out-Of-Hospital Trial

    Van Alem AP, Chapman FW, Lank P, Hart AAM, Koster RW. A prospective, randomised and blinded comparison of first shock success of monophasic and biphasic waveforms in out-of-hospital cardiac arrest. Resuscitation 2003;58(1):17- LIFEPAK 1000 Defibrillator Operating Instructions To Purchase, Visit Avobus.com...
  • Page 93: G Electromagnetic Compatibility Guidance

    APPENDIX G ELECTROMAGNETIC COMPATIBILITY GUIDANCE LIFEPAK 1000 Defibrillator Operating Instructions ©2018 Physio-Control, Inc.
  • Page 94 To Purchase, Visit Avobus.com or call 1-800-674-3655...
  • Page 95 The LIFEPAK 1000 defibrillator is intended for use in the electromagnetic environment specified below. The customer or the user of the LIFEPAK 1000 defibrillator should ensure that the device is used in such an environment. Guidance and Manufacturer’s Declaration - Electromagnetic Emissions...
  • Page 96 Tables 2 through 4. The LIFEPAK 1000 defibrillator is intended for use in the electromagnetic environment specified below. The customer or the user of the LIFEPAK 1000 defibrillator should ensure that the device is used in such an environment.
  • Page 97 Electromagnetic Compatibility Guidance The LIFEPAK 1000 defibrillator is intended for use in the electromagnetic environment specified below. The customer or the user of the LIFEPAK 1000 defibrillator should ensure that the device is used in such an electromagnetic environment. Guidance and Manufacturer’s Declaration - Electromagnetic Immunity...
  • Page 98 Electromagnetic Compatibility Guidance SEPARATION DISTANCES The LIFEPAK 1000 defibrillator is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the LIFEPAK 1000 defibrillator can help prevent electromagnetic interference by maintaining a minimum distance...
  • Page 99 (m) can be determined using the equation in the table above with the known power, P of the transmitter in watts (W) according to the transmitter manufacturer. Note 1:These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. LIFEPAK 1000 Defibrillator Operating Instructions ©2018 Physio-Control, Inc.
  • Page 100 To Purchase, Visit Avobus.com or call 1-800-674-3655...
  • Page 101 Recycling 5-8 cprMAX Technology viii defibrillator 5-1 Defibrillator data 4-2 Daily self-test viii Caution 1-2 Deleting of patient data 4-2 Data management viii Device ID screen D-7 ECG display (optional) vii Cautions LIFEPAK 1000 Defibrillator Operating Instructions Index-1 ©2018 Physio-Control, Inc.
  • Page 102 Warranty information 5-9 Pacemakers, patients with implanted 3-5 Packaging recycling 5-8 Pediatric patients 3-3 Physical Characteristics A-7 Placing electrode pads 3-3 Product recycling 5-8 Pulse setup submenu D-5 Readiness Display viii Rechargeable battery 5-5 Recycling 5-8 Index-2 LIFEPAK 1000 Defibrillator Operating Instructions...
  • Page 103 To Purchase, Visit Avobus.com or call 1-800-674-3655...
  • Page 104 Physio-Control is now part of Stryker. For further information, please contact your local Physio-Control representative or visit www.physio-control.com. Physio-Control, Inc. Stryker Australia Pty Ltd 11811 Willows Road NE 8 Herbert Street Redmond, WA 98052 USA St Leonards Telephone: 425.867.4000 NSW 2065 Fax: 425.867.4121 Australia www.physio-control.com...