6.1.2. Device parameters for the assignment of Pt100(0), thermocouple: Pt.SE, THEr
Pt.SE: Three variants are available:
Pt.Lo: Pt100 3-wire -50.0...200.0°C / -58.0...392.0°F
Pt.Hi: Pt100 3-wire -200...850°C / -328...1562°F
Pt.tH: Pt1000 2-wire -200...850°C / -328...1562°F
tHEr: Here we differentiate between : Thermocouple type L, J, K, B, S, N, E, T, R
Parameter
Pt.SE
Parameter
tHEr
Parameter
UnIt
OFFS
UnIt: Type of temperature measurement.
With Unit you select the representation of the temperature in °C or °F.
OFFS: Line adjustment.
The value adjustment for a temperature measurement in °C can be set between -20.0 and +20.0
and for a later measurement in °F between -36 and +36. If the type of measurement is switched
later, the value is rounded.
General display parameters see page 27
Alarm parameters see page 28
Backup parameters for locking / unlocking the parameterization see page 30
Selection option
Selection option
Selection option
to
to
19
6. Parameterization
Default
Default
Default