6. Parameterization
6.1. Input signal selection: type
With the type setting an assignment of the input variant takes place, here can be selected
between the five input types voltage, current, Pt100(0), thermocouple and impulse signal.
Measurement signal
Measuring signal
Alarm parameters
selection
Special
parameters
General display
parameters
Safety
parameters
Measuring time, moving averaging, start/end
value representation in the display, key function,
display flashing
see page 27
Limit value behavior, limit value message,
switching behavior, switching threshold,
hysteresis, delay
see page 28-30
Locking / releasing the parameterization
see page 30
15
6. Parameterization
Voltage
see page 15-18
Power
see page 15-18
Pt100(0)
see page 19
Thermocouple
see page 19
Impulse input
see page 20-22