OPTILAB TWL-C-M Series User Manual page 15

Tunable wavelength laser, c-band module
Table of Contents

Advertisement

TWL-C-M User's Manual
C
OMMAND
① S
M
CAN
ODE
S
ETTING
② E
/D
NABLE
ISABLE
S
M
CAN
ODE
③ D
S
EVICE
TATUS
Optilab, LLC
600 E. Camelback Road, Phoenix, AZ 85012
Phone: (602) 343-1496, Fax: (602) 343-1489, Email: sales@oequest.com
On 'Scan Setting', these boxes are for entering/selecting in the
desired scan mode values.
First, user needs to enter 'Serial Number' on text field. User
can find such information on module or test report.
[Note that selecting wrong 'Serial Number' and proceeding
scan function may cause laser damage of TWL-C-M unit]
'Scan Range' needs to be entered by user. It provides two
options, 'Entire Scan Range' or 'Manual Range Setting'. If
'Entire Scan Range' box has checked, scanning range will be
set from first to last available frequency of unit. If checkbox
remains unchecked, user needs to enter desired range of
scanning. Available range can be observed from the 'Scan
Range : " or 'General' tab.
'Sweep Range' needs to be entered by user within the range of
'50 ~ 250' in terms of GHz. This range defines center
frequency of one sweep among total sweeps (Clean Scan).
[We recommend 100 GHz for optimal speed and coverage]
'Target Power' needs to be selected by user from dropdown
menu. It is limited to 10 dBm (Optilab TWL GUI v1.0).
'Scan' button will initiate scanning of laser. Once it starts
scanning, you can stop scan function at any point with 'Stop'
button
[Once 'Stop' button has clicked, please allow '~ 3 second' to
fully stop]
This box will display unit's 'Center Frequency' of sweep and
current status of scanning
'Current Center Frequency' will display center frequency of
current sweep range (range defined by 'Sweep Range') in
terms of 'THz'
'Status box' tells user the current status of scanning such as
start, cancel or stop. User can delete record with 'Clear' button
D
ESCRIPTION
Page 12 of 16