Electrical Power Source - Universal Laser Systems VLS2.30 Setup Manual

Hide thumbs Also See for VLS2.30:
Table of Contents

Advertisement

Installation & Set-up Guide
Electrical Power Source (User Supplied)
1. For your system's electrical requirements, please refer to the "INPUT POWER" label near the
power inlet.
CAUTION:
2.
into a non-grounded outlet. A laser system that is not properly grounded is hazardous and
has the potential to cause severe or fatal electrical shock. Without proper grounding, the laser
system may exhibit sporadic or unpredictable behavior. Always plug the system into a properly
grounded (earthed) outlet.
3. Noisy or unstable electricity and voltage spikes may cause interference and possible damage
to the electronics of the laser system. If electrical power fluctuations, brown outs or constant
power outages are a problem in your area, please contact your local Electrician to supply a
power isolation and regulation module. Electrically noisy equipment, such as equipment with
large motors, can also cause interference if plugged into the same outlet. It may be necessary
to attach the laser system to a dedicated electrical line to resolve the problem.
4. The laser system is designed as a Class I, Group A, pluggable device. It is also designed for
connection to IT Power systems which provide the most flexibility to the user.
6
Never remove the ground lead to the electrical cord and plug the laser system