The Diamond is suitable for use in the specified electromagnetic environment. The purchaser or user
of the Diamond should assure that it is used in an electromagnetic environment as described below:
Immunity Test
Voltage dips,
short
interruptions and
voltage
variations on
power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
n
Table 10.2 Electromagnetic immunity IEC 60601-1-2
86500-IMG rev 3
IEC 60601-1-2
Test Level
0 % U
for 5 sec
0 % U
n
3 A/m
Compliance
Electromagnetic Environment
Level
for 5 sec
Mains power quality is that of a
n
typical commercial and/or hospital
environment. If the user of the
Diamond requires continued
operation during power mains
interruptions, it is recommended that
the Diamond be powered from an
uninterruptible power supply or a
battery.
3 A/m
Power frequency magnetic fields are
at levels characteristic of a typical
location in a typical commercial and/
or hospital environment.
GE Healthcare
10 Specifications
79