10 Specifications
System specifications for Diamond, model MGX 2000
IEC applied part type
Enclosure
Weight
Height
Operation mode
Radiation intensity
Environmental specifications
Operating conditions
Storage and transportation
Humidity
Pressure
Manufacturer
Type
Frequency
Power
kV range
Accuracy
kV ripple
mA range
Maximum mA output
72
B
IP 20
350 kg (771 lbs.)
1.9 m (6 1/4 ft)
Continuous
operation
intermittent loading
Typical range 1000-1200mR/
100mA at 28kV
+10°C...+40°C
Non condensing (maximum
humidity 90%)
(+0°C...+35°C)
10%...100%
500 mbar - 1060 mbar
Generator
GE Healthcare Finland Oy
MGHV-300
40 - 100 kHz
Max. 2.6 kW
15 - 39 kV, in 1 kV increments
± 1.5 kV
< 4%
Focus 0.3 mm 50 - 100 mA
Focus 0.3 mm 25 - 50 mA (low mA
mode)
Focus 0.1 mm 18 - 30 mA
(- tube current reduced with small
breast thickness)
Focus 0.3 mm 150 mA
Focus 0.1 mm 45 mA
GE Healthcare
with
86500-IMG rev 3