Emergency information for all LaMotte reagents is available from Chem-Tel
(US, 1-800-255-3924) (International, call collect, 813-248-0585).
These color standards are to be used as Chlorine Secondary Standards and are not
intended to be used as calibration standards.
If the color standards are to be used with more than one colorimeter, copy and fi ll
out the Data Sheet for each meter. Do not write the standard values on the color
standard caps.
These color standards are to be used as Chlorine Secondary Standards to verify the
performance of the meter and are not intended to be used as calibration standards.
The meter is calibrated at the time of manufacture. The standards should be read on
the meter when they are received and the results should be recorded. Periodically, as
determined by the user, the standards should be read on the same meter and the
results should be compared to the initial readings. The meter is performing satisfactorily
if the difference between the initial readings and the current readings is within the
specifi ed tolerances on the data sheet.
The initial readings for the standards should be approximately:
Standard A
Standard B
Standard C
The exact value of the standards will vary on different types of meters. The critical
measurement is the change in the readings over time, not the value of readings.
COLORIMETER
CHLORINE SECONDARY STANDARDS
CODE 4140-03
CONTENTS
DPD Chlorine Color Standard Blank
DPD Chlorine Color Standard A
DPD Chlorine Color Standard B
DPD Chlorine Color Standard C
0.1 - 0.4 ppm
0.9 - 1.2 ppm
2.2 - 3.0 ppm
CODE
4136-E
4147-E
4138-E
4139-E

Advertisement

loading

Summary of Contents for LaMotte 4140-03

  • Page 1 4138-E DPD Chlorine Color Standard C 4139-E Emergency information for all LaMotte reagents is available from Chem-Tel (US, 1-800-255-3924) (International, call collect, 813-248-0585). These color standards are to be used as Chlorine Secondary Standards and are not intended to be used as calibration standards.
  • Page 2 Procedure in the Colorimeter manual. Use two standards, as described, that have been titrated by an EPA accepted method (such as LaMotte Wipe the DPD Chlorine Color Standard Method 3176-02). Blank (4136-E) with a clean, lint-free cloth. Push the READ Insert the...
  • Page 3 DC1200 - INSTRUMENT PERFORMANCE CHECK Perform this procedure at specifi ed intervals to verify instrument performance. Insert the DPD Chlorine Color Standard Blank into the chamber, being sure to align the index line with the arrow on the meter. Close the lid. Wipe the DPD Chlorine Color Standard Blank (4136-E) with a clean, lint-free cloth.
  • Page 4 Procedure in the Colorimeter manual. Use two standards, as described, that have been titrated by an EPA accepted method (such as LaMotte Wipe the DPD Chlorine Color Standard Method 3176-02). Blank (4136-E) with a clean, lint-free cloth. Insert the DPD Chlorine Press ON to turn the meter on.
  • Page 5 SMART 2 - INSTRUMENT PERFORMANCE CHECK Perform this procedure at specifi ed intervals to verify instrument performance. Insert the DPD Chlorine Color Standard Blank into the chamber, being sure to align the index line with the arrow on Wipe the DPD Chlorine Color Standard the meter.
  • Page 6 Procedure in the Colorimeter manual. Use one standard, as described, that has been titrated by an EPA accepted method (such as LaMotte Wipe the DPD Chlorine Color Standard Method 3176-02). Blank (4136-E) with a clean, lint-free cloth. Insert the DPD Chlorine Press to turn the meter on.
  • Page 7 SMART3 - INSTRUMENT PERFORMANCE CHECK Perform this procedure at specifi ed intervals to verify instrument performance. Insert the DPD Chlorine Color Standard Blank into the chamber, being sure to align the index line with the arrow on the meter. Close the lid. Wipe the DPD Chlorine Color Standard Blank (4136-E) with a clean, lint-free cloth.
  • Page 8 Procedure in the Colorimeter manual. Use one standard, as described, that has been titrated by an EPA accepted method (such as LaMotte Wipe the DPD Chlorine Color Standard Method 3176-02). Blank (4136-E) with a clean, lint-free cloth. Insert the DPD Chlorine Press to turn the meter on.
  • Page 9 1500 - INSTRUMENT PERFORMANCE CHECK Perform this procedure at specifi ed intervals to verify instrument performance. Insert the DPD Chlorine Color Standard Blank into the chamber, being sure to align the index line with the arrow on the meter. Close the lid. Wipe the DPD Chlorine Color Standard Blank (4136-E) with a clean, lint-free cloth.
  • Page 10: Data Sheet

    DPD. Readings of actual DPD chlorine reactions will not differ signifi cantly between meters. Contact LaMotte Company Technical Service at 1-800-344-3100 or tech@lamotte.com with questions. Serial Number:...
  • Page 11 BLANK...
  • Page 12 LaMOTTE COMPANY Helping People Solve Analytical Challenges PO Box 329 • Chestertown • Maryland • 21620 • USA 800-344-3100 • 410-778-3100 (Outside U.S.A.) • Fax 410-778-6394 Visit us on the web at www.lamotte.com 64140-03 1/16...