Table 202: Guidance and manufacturer's declaration
®
The NeuroTrac
environment specifi ed below. The customer or the user of the NeuroTrac
product should assure that it is used in such an environment, and that
precautions regarding that environment are heeded.
Immunity test
Electrostatic
discharge
(ESD)
IEC 61000-4-2
Power fre-
quency (50/60
Hz)
magnetic fi eld
IEC 61000-4-8
– electromagnetic immunity
product is intended for use in the electromagnetic
IEC 60601
test level
6 kV contact
±8 kV air
3 A/m
®
NeuroTrac
Sports XL Operation Manual
Compliance
Electromagnetic environ-
level
Floors should be wood,
concrete or ceramic tile.
±6 kV con-
If fl oors are covered with
tact
synthetic material, the rela-
±8 kV air
tive humidity should be at
least 30 %.
Power frequency magnetic
fi elds should be at levels
characteristic of a typical
3 A/m
location in a typical com-
mercial or hospital environ-
ment.