Setting up Customised Programmes PC1, PC2 or PC3
First, if a programme is running, press the PRG button twice to return to the
home screen.
Refer to the example of custom programme diagram on page 9.
1.
Press the PRG button until PC1, PC2 or PC3 is selected.
2.
Press and hold the SET button for three seconds, the phase indicator
arrow and Hz symbol will  ash on and off .
3.
Press CH.B +/- to set the frequency between 2 Hz and 100 Hz.
4.
Press CH.A +, the μS symbol will  ash, press CH.B +/- to set the
pulse duration between 50 μS and 330 μS.
5.
Press CH.A +, the MIN symbol will  ash, press CH.B +/- to set
the length of the phase time between 1 and 99 minutes.
Set the time to zero to end the programme on this phase.
6.
Press CH.A +, the WORK / REST or the CONT symbols will  ash,
Press the CH.B +/- to select WORK / REST or CONT (continuous).
Note: if continuous is selected, the menu will loop back to step 2.
7.
Press CH.A +, WK will appear and  ash, press CH.B +/- to set the
work seconds between 2 and 99 seconds.
8.
Press CH.A +, RT will appear and  ash, press CH.B +/- to set the rest
seconds between 2 and 99 seconds.
9.
Press CH.A +, RP will appear and  ash, press CH.B +/- to set the ramp
seconds between 0.1 and 9.9 seconds.
10.
Press CH.A +, AL or SY will appear and  ash, press CH.B +/- to
select alternating or synchronous current.
Note: if alternating is selected, the menu will loop back to step 2.
11.
Press CH.A +, DY will appear and  ash, press CH.B +/- to set the
delay of channel B starting between 0 and 4 seconds after channel A.
12.
The menu will now loop back to step 2 and the Hz symbol will  ash.
13.
To set the next phase, press the set button. The phase symbol will  ash
over the next phase, continue with step 2 to set this phase.
14.
When  nished setting the phases, press the PRG button to save the
settings and return to the home screen.
The programme will be saved permanently.
Setting the phase time of phase 2,3,4 or 5 to zero will cause the programme to
end at that phase.
Following procedures 1 to 12 can reprogramme a customised programme.
If for example there are 5 pre-set phases in one overall programme and only
4 phases are now required, input 0 (zero) into the phase time that is no longer
required and press the PRG key to store the new information.
®