Preflight Checkout - Agilent Technologies 8510C Service Quick Reference Manual

Network analyzer system
Hide thumbs Also See for 8510C:
Table of Contents

Advertisement

Be sure security key is installed
on A8 board for Network Analyzer
Std., Opt. 010, or Antenna software.
Note: Opt. 010 also requires firmware
key.
Security Keys
for 85101C Security Board A8
85101-69268 < Opt. 010 85101-80091
08530-69001 < Antenna 08530-80004
CRT Only
85101-69273 < Rev. 6.XX – Rev. 8.XX
85101-80114
Rev. C.06.60 85101-80132 CRT Only
Rev. C.07.XX 85101-80116 CRT Only
Rev. C.08.XX 85101-80116 LCD Only
Reference extensions A and B are
used to phase balance a sweeper to
make an uncalibrated measurement.
At the port use 08512-20019 short.
With a cable(s) use 08514-20013 long.
If in doubt, always use the short links.
83621A & 51A or others need
16 Aug 91 or later to work with
performance verification or else
unable to match frequency
points messages.
Firmware Rev. March 8, 1991
for complete compatibility
(If not, you may lose quick step)
8516A is compatible with 8510B
Rev. 4.0 or higher and with
8340/41/A/B with Rev. May 11, 88
or later. If not use Upgrage Kit 11875A
for 8340/41/A/B with serial numbers
2513 and above.
Use Upgrade Kit 08340-60334 for
serial numbers 2504A and below.
8517B Opt. 007 will display spikes if
the amplifiers are overdriven from the
RF Source. Reduce RF power to
0 dBm to increase dynamic range,
You may need to raise the power
> than 0 dBm. The factory preset
puts it at +10 dBm.
NOTE:
-20 dBm will cause "NO IF FOUND".
SWEEP IN 0-10V
10MHz REF IN
To 85102
SWEEP OUTPUT
SOURCE CONTROL
STOP SWEEP
Exceptions andNo tes:
1. A 9000 series 200 or 300 computer with 2.5 MB of memory
is recommended; not 9826A. 9816 limited to 2 megabytes.
2. Connect the reference port extension cables as shown above.
(Not applicable for 85110, 8516, 8517.)
3. Only use factory preset (not user preset) during service. Be sure
multi-source, phase lock external or phase lock none are turned off.
8510C Installation

"Preflight Checkout"

PRINTER
(OPTIONAL)
ADDRESS 1
PLOTTER
ADDRESS 5
9153C
OPT 010/020
9122C
DISK DRIVE
(ADDRESS 0)
85101
DISPLAY/
PROCESSOR
(ADDRESS 16)
IF DISPLAY INTERCONNECT
85102
IF DETECTOR
(ADDRESS 16)
851X A/B
TEST SET
(ADDRESS 20)
X = 1-7
RF OUTPUT
8360 SERIES
RF SOURCE
(ADDRESS 19)
With 8510B must
have Rev. 4.0 or
Higher to work
properly with
8360 Sources.
**
Source Control Cable
used with 8516A only
*
Note:
For 8340 or 8350, a BNC open must be
connected to the 85102 trigger input.
For 8360, a BNC open 1250-0053 should be
connected
*
*
Note:
8340
Cannot be mixed
SOURCE
with 8360s.
(ADDRESS 19)
RF INPUT
To Test Set
RF OUTPUT
10MHz
OUTPUT
JUMPER
GPIB SYSTEM BUS
To 85102
To 85101
PRINTERS
PLOTTERS
2225A
33447A
7440A
2225D
33440A
7475A
D2106A
33471A
7550A
3630A
33449A
7550B
System Bus
Ext Display
GPIB
*
*
EXTENSION A
EXTENSION B
RF INPUT
Source Interface
Source Control
**
10MHz
Auxilliary
Interface
to the 85102 SWEEP IN 0-10V.
(ADDRESS 19)
RF INPUT
To Test Set
Firmware Rev 6 for 8350B
RF OUTPUT
Note:
Some 83592A's do not have
DC block cap so will not work
with 8510=no phase lock.
To upgrade the 8350A to work with the 8510, order upgrade kit
Part No. 08350-60100
4. An 8340 or 8360 requires a source interconnect cable
(Part No. 08510-60009) connection when the test set is
used as an 8516. 8350 cannot be used with 8516 above 20 GHz.
5. Instrument power-on sequence: Source, Test Set, 85102, 85101.
Check power line modules for local line voltage. Check GPIB
addresses of all instruments. For 8360, set language switch to 001.
For 8340/41 set FREQUENCY STANDARD switch to INTERNAL.
For a complete list
of compatible
printers/plotters
see the 8510C
Ordering Guide.
GPIB Cables
8120-3445
Ecal requires Rev. 4.0
External
Controller
Master Calib. Constants
disk 08510-10034 A.6
STOP SWEEP
Never used with
SWEEP IN 0-10V
8360 Series
J10 TEST SET INTERCONNECT
10MHz IN
TRIGGER INPUT
J11 TEST SET INTERCONNECT
TRIGGER OUTPUT
STOP SWEEP
If connected to 85102B:
SWEEP OUTPUT
8510C system will lock up
or not work properly.
DO NOT USE WITH 85102B/8510C!
001
19
000 = TMSL/SCPI
1
010 = MATE LANG
0
001 = NA LANG ADDRESS = 19
L L L
A A A A A
111 = OVERIDE
3
2
1
5
4
3
2
1
You will need to cycle power to accept new settings.
8350B
SOURCE
Firmware rev
6 for 8350B main frame - press [SHIFT] [49]
Firmware rev
7 for 8359X series plug ins - press [SHIFT] [99]
Firmware rev
6 for 835XX series plug ins - press [SHIFT] [99]
SWEEP OUTPUT
SWEEP OUT/IN
To 85101
System Bus
RF INTERCONNECT
minimum
BNC Cables
8120-1840
Not used with
8510 A/B
Rev. 5.14 or less.
OK for8510B/C
6.0 or greater.
To 85102
STOP SWEEP
SWEEP IN 0-10V
rs411c

Advertisement