Performance Verification
Optional Swept Frequency Accuracy Check
Table 4-4 Performance Test Record for Optional Test
Instrument Model
______________
Test Description
Swept Frequency Accuracy
Start Frequency:
________________
Stop Frequency:
________________
The measurement uncertainty is quoted for these performance tests using only the
a.
recommended models specified in Table 8-1 of the On-Site Service Manual. The quoted
uncertainty represents limits of ± 3 times the equivalent standard deviation (3 σ) and is
intended to represent a 90% confidence level.
In Case of Difficulty
If the measured values do not meet the specifications listed on the test record, refer to your
source manual for adjustment and troubleshooting instructions.
4-12
Report Number
____________
Recorded Results
Maximum Specification
_____________
0.1% of sweep
(for 8360)
_____________
1% of sweep
Date
___________
Measurement Uncertainty
± 150 kHz
± 150 kHz
8510C Service Quick Reference Guide