Figure 4-3
Full Span Response at 1 MHz per Division
For 8340/41 Series Synthesizers
Position the highest frequency bandswitch transition point on to the reference line. Press
synthesizer keys:
CAL, [MORE], [TRIM SWEEP]
trace variation on the test record.
Figure 4-4
Trace Variation Display
8510C Service Quick Reference Guide
and then adjust the front panel knob. Record the maximum
Figure 4-4
shows a typical display.
Performance Verification
Optional Swept Frequency Accuracy Check
4-11