Dedicated 8510 System Source Models - Agilent Technologies 8510C Service Quick Reference Manual

Network analyzer system
Hide thumbs Also See for 8510C:
Table of Contents

Advertisement

Dedicated 8510 System Source Models

Dedicated sources are optimized for use as 8510 system components. They are configured
without modulation capabilities or front panel keyboard/displays. They have rear
connectors, and one-year on-site service where available. Specifications for these models
are the 8510 specifications with the following additions:
Frequency Range (all serial prefixes)
83621A/B
83631A/B
83651A/B
Resolution
Maximum Leveled Output Power
(all serial prefixes)
Frequencies ≤ 20 GHz
Frequencies > 20 GHz and ≤ 26.5 GHz
Frequencies > 26.5 GHz and ≤ 40 GHz
Frequencies > 40 GHz
Minimum Settable
If your source does not meet these power requirements, you may need to retrack the YIG
oscillator. When you look at power, also check for spikes, powerholes, and squegging.
8510C Service Quick Reference Guide
Dedicated 8510 System Source Models
45 MHz to 20 GHz
45 MHz to 26.5 GHz
45 MHz to 50 GHz
1 Hz
+ 10 dBm
+ 4 dBm
+ 3 dBm
0 dBm
− 20 dBm
Instrument Compatibility
2-5

Advertisement

Table of Contents
loading

Table of Contents