Beamage-M² User Manual
Revision 6
28
If the quadratic fit curve passes under zero, the PC-Beamage software will try to solve the problem by
deleting the measurements farthest from 0 and by automatically recalculating the value of M². The PC-
Beamage software will also offer a recommended travel distance between −3.5 and 3.5 to correct the
problem (see Figure 5-25).
Figure 5-25: Curve Fit Passes Under Zero.
A minimum of 5 beams must be detected to calculate M
2
.
Upon clicking the START ISO SCAN button, click Yes if you want to save previous data or No to ignore
them. The data erase themselves and then a new M² measurement begins automatically.
Figure 5-26: Save previous data
Once the new measurement is completed, click the Calculate M² button to display the M² Results tab.