Figure 5-25: Curve Fit Passes Under Zero; Figure 5-26: Save Previous Data - Gentec Beamage M2 User Manual

Table of Contents

Advertisement

Beamage-M² User Manual
Revision 6
28
If the quadratic fit curve passes under zero, the PC-Beamage software will try to solve the problem by
deleting the measurements farthest from ��0 and by automatically recalculating the value of M². The PC-
Beamage software will also offer a recommended travel distance between −3.5���� and 3.5���� to correct the
problem (see Figure 5-25).

Figure 5-25: Curve Fit Passes Under Zero.

A minimum of 5 beams must be detected to calculate M
2
.
Upon clicking the START ISO SCAN button, click Yes if you want to save previous data or No to ignore
them. The data erase themselves and then a new M² measurement begins automatically.

Figure 5-26: Save previous data

Once the new measurement is completed, click the Calculate M² button to display the M² Results tab.

Advertisement

Table of Contents
loading

Related Products for Gentec Beamage M2

This manual is also suitable for:

203605