Summary of Contents for Koehler K60094
Page 1
Portable Heated Oil Test Centrifuge Operation and Instruction Manual REV B Koehler Instrument Company, Inc. 1595 Sycamore Avenue • Bohemia, New York 11716-1796 • USA Toll Free: 1-800-878-9070 (US only) • Tel: +1 631 589 3800 • Fax: +1 631 589 3815 http://www.koehlerinstrument.com •...
Page 3
CERTIFICATE OF CONFORMANCE Heated Oil Test Centrifuge K60094 This certificate verifies that part number K60094, Heated Oil Test Centrifuge, was manufactured in conformance with the applicable standards set forth in this certification. Specifications: ASTM D96 API MPMS Chapter 10.4 API 2542 This unit is tested before it leaves the factory, to ensure total functionality and compliance to the above specifications and ASTM standards.
Dimensions: 20.79 x 23.43 x 13.91 the best solutions. You can depend on Koehler for a (w x d x h, in.(cm)) (with pre-heater lid open) full range of accurate and reliable instrumentation as (52.8 x 59.5 x 35.33)
Packing List NEVER leave the centrifuge unattended while unit is on. K60094 Portable Heated Oil Test Centrifuge, 12VDC 40A Ensure lid switch is in locked position prior to Short Cone Centrifuge Tube Holder (2) spinning rotor.
K60094 Portable Heated Oil Test Centrifuge Operation and Technical Manual return goods to Koehler without written authorization. 4 Descriptions Portable Oil Test Centrifuge Installation Equipment Placement: Place the instrument on a firm surface in a secure area. The equipment comes with the z-brackets installed in the factory.
K60094 Portable Heated Oil Test Centrifuge Operation and Technical Manual 5. Pre-heater Power. Controls power to the two Control Panel preheating chambers. These chambers are the two right most chambers of the integrated four place tube holder assembly. Press switch to the right to power on.
K60094 Portable Heated Oil Test Centrifuge Operation and Technical Manual 4. Down Arrow. Used to decrease the set centrifuge tubes to exactly the 50-mL point temperature and to decrease or (100-part) mark with the test sample. change parameters when programming temperature controller.
Page 11
K60094 Portable Heated Oil Test Centrifuge Operation and Technical Manual 15. Read and record the combined volume of Make sure there are cushions at the bottom sediment and water at the bottom of each of the tube holders and there is no oil tube.
K60094 Portable Heated Oil Test Centrifuge Operation and Technical Manual 6 Maintenance Replacement Parts Some instrument parts may need to be replaced. WARNING. Disconnect power to the unit before When ordering replacement part(s), please provide servicing to avoid exposure to high voltages and/or...
K60094 Portable Heated Oil Test Centrifuge Operation and Technical Manual Troubleshooting 11 Warranty 8.1 The test will not start We at Koehler would like to thank you for your equipment purchase, which is protected by the Contact Koehler’s Technical following warranty.
Need help?