11.3.2 Cradle cleaning and disinfection

Cleaning
1. Using CaviWipes1, remove any gross contaminants on the cradle for a minimum of one (1) minute.
Note: If you are using an alternative disinfectant, please refer to
Approved cleaning and disinfecting materials
for the required contact time.
Figure 206: iTero Element 5D – Remove gross contaminants on the cradle using CaviWipes1
Figure 207: iTero Element 5D Plus cart – Remove gross contaminants on the cradle using CaviWipes1