Calibration; Maintenance And Repair; Environmental Requirements - PCE Health and Fitness PCE-CT 100N-ICA User Manual

Coating thickness gauge
Table of Contents

Advertisement

8.5.3
File processing
Press the Menu button to enter the Instrument Menu Settings screen.
Use the navigational buttons to navigate the menu to the data icon.
Press the menu button (≡) to enter the data settings.
Select the File No. setting and use the ◄ or ► buttons to change the file set number.
The device supports up to 5 file sets.
Select the File Browser and use the ▼ to ▲ buttons to navigate the 5 file sets. Use the
menu button (≡) to view a file set. Each file set can contain up to 28 data sets.
Select File Clear and use the menu button (≡) to confirm the process. The entire
selected file set will be cleared.
9

Calibration

The device and impact probe must be calibrated with a coating thickness matrix before first use
or after a prolonged period of non-use.
Press the Menu button to enter the Instrument Menu Settings screen.
Use the navigational buttons to navigate the menu to the calibrate icon.
Press the menu button (≡) to enter the data settings.
6 measurements must be performed vertically down into the provided substrate; the first is a zero
calibration (without the substrate).
Use the ▲ or ▼ buttons to adjust the value to the nominal thickness of the standard
test substrate for each measurement.
Perform a total of 5 measurements; one test for one test foil.
Press the menu button (≡) to save and exit the calibration process.
10

Maintenance and Repair

10.1

Environmental requirements

Strictly avoid collisions, heavy dust, moisture, strong magnetic fields, oil stains, and other dirtying