5.5 Typical M Factor Temperature Dependency 5.6 Drawings 5.7 Fiber Coupling onto Small Detector Area 5.8 Certifications and Compliances 5.9 Warranty 5.10 Copyright and Exclusion of Reliability 5.11 Thorlabs 'End of Life' Policy 5.12 List of Acronyms and References 5.13 Thorlabs Worldwide Contacts...
If necessary, ask for replacement pack- aging. Refer servicing to qualified personnel! Only with written consent from Thorlabs may changes to single components be made or components not supplied by Thorlabs be used. All modules must only be operated with duly shielded connection cables.
Note If you want to use your own power supply, you can ask Thorlabs for an appropriate power con- nector cable. · Carefully unpack the unit and accessories. If any damage is noticed, do not use the unit.
For fiber coupled applications, in case of detectors with an active area of 1 mm, fiber adapters like Thorlabs S120-xx series can be easily mounted on the optical input. The fiber adapter will accommodate multi-mode as well as single-mode fiber. For detectors with smaller active areas, in order to meet specifications, it is recommended to focus the optical signal out of the fiber onto the detector.
The optical damage threshold is 1 mW. Exceeding this value will permanently destroy the Avalanche Photodetector! 3.1.2 Electrical Output Thorlabs APD110x/120x Avalanche Photodetectors deliver an OUTPUT voltage, which is a function of incident light power , detector's responsivity (l) at a given wavelength and M Â...
The slim line housing comes with a removable threaded coupler that is compatible with any number of Thorlabs 1” and ½” threaded accessories. This allows convenient mounting of ex- ternal optics, filters, apertures or fiber adapters, as well as providing an easy mounting mechan- ism using the Thorlabs cage assembly accessories.
Another critical point can be electrostatic coupling of electrical noise associated with ground loops. In most cases an electrically isolated post (see Thorlabs parts TRE or TRE/M) will sup- press electrical noise coupling. You should always try to identify the electrical noise sources and increase the distance to the Avalanche Photodetector.
The unit does not need a regular maintenance by the user. It does not contain any modules and/or components that could be repaired by the user himself. If a malfunction occurs, please contact Thorlabs for return instructions. Do not remove covers! High voltage!
For detectors with an active area diameter of 1.0 mm, a Thorlabs S120-xx fiber adapter is re- commended. In this case, the detector area will not be overfilled by the diverging beam out of the fiber tip.
For warranty repairs or service the unit must be sent back to Thorlabs. The customer will carry the shipping costs to Thorlabs, in case of warranty repairs Thorlabs will carry the shipping costs back to the customer.
Waste treatment on your own responsibility If you do not return an “end of life” unit to Thorlabs, you must hand it to a company specialized in waste recovery. Do not dispose of the unit in a litter bin or at a public waste disposal site.
Need help?
Do you have a question about the APD110 Series and is the answer not in the manual?
Questions and answers