Short-term pressure constant
measured according to
prEN 17510:2005 and German
HMV criteria in CPAP mode
Long-term pressure constant
measured according to
prEN 17510:2005
Finefilter separation level
to 2 µm
Fine filter service life
Permitted humidity for operation
and storage
System resistance with an air flow
of 60 l/min at the patient
connector aperture.
All values determined under ATPD conditions (ambient temperature and pressure,
dry).
BiLevel ST 22
at
at
at
at
at
at
at
∆p = 0.2 hPa
1000 hours with normal room air
≤ 95 % rF (no condensation)
BiLevel ST 22 with hose system
WM 24130 and Silentflow
WM 23600
kPa · s
0,20
l
BiLevel ST 22 with VENTIclick
20 hPa: ∆p ≤ 1 hPa
15 hPa: ∆p ≤ 1 hPa
14 hPa: ∆p ≤ 1 hPa
10 hPa: ∆p ≤ 1 hPa
7 hPa: ∆p ≤ 0.5 hPa
5 hPa: ∆p ≤ 0.5 hPa
4 hPa: ∆p ≤ 0.5 hPa
≥ 99.7%
BiLevel ST 22 with O
system WM 23737, VENTIclick
WM 24365 and bacteria filter
WM 24148
0,31
Technical data
hose
2
kPa · s
l
49