5.2 Cleaning

Cleaning the hose system
1. Remove the hose system from the unit and exhalation system.
3. Using a small amount of detergent in warm water, clean the corrugated tube so
no residues remain. Rinse the inside of the hose thoroughly.
4. Rinse the corrugated tube thoroughly both inside and out using clean, warm
water.
5. Shake the hose system thoroughly.
6. Hang up the hose system and allow it to drip to prevent moisture from getting
into the BiLevel ST 22.
7. Remove the plugs from the pressure measurement hose.
Dry the hose system using BiLevel ST 22
26
Hygienic preparation
2. Pull one end of the pressure measurement hose out
(shake gently if necessary) and seal it with the
sealing plug supplied. At the other end, seal off
the small opening of the adapter with the second
sealing plug so that no water can enter.
1. If water should accidentally get into the pressure
measurement hose, put the supplied red drying
adapter into the unit output.
2. Push the adapter of the hose system onto the red
drying adapter.
3. Press the On/Off switch
process. Use the menu key
patient menu. Press the dial to start the drying
process. The unit will now dry the hose system.
to start the drying
to access the