Bacteria filter
When using the bacteria filter WM 24148 , change the particle filter in the bacteria
filter in accordance with the relevant Operating Instructions.

8.3 Change the pressure measurement hose

Adapter
Sleeve
Pressure measurement hose
2. The "Reset filter change?" prompt appears. Using
the dial, select
dial.
To cancel the process, select
dial. The process will be canceled.
If you have selected and confirmed
message "Filter Change reset" appears for 3
seconds.
1. Remove the sleeve of the corrugated tube from the
adapter.
2. Pull the pressure measurement hose out of the
corrugated tube.
3. Pull the pressure measurement hose away from the
adapter.
4. Put the new pressure measurement hose onto the
adapter.
5. Hold the corrugated tube up and introduce the
free end of the new pressure measurement hose.
6. Push the sleeve of the corrugated tube onto the
adapter.
7. Dispose of the old pressure measurement hose.
and confirm by pressing the
YES
NO
Maintenance
and press the
, the
YES
41