⚫ Do not allow liquid to enter the case.
⚫ Never use abrasive material (such as steel wool or silver polish).
⚫ Inspect the system and reusable accessories after they are cleaned and disinfected.
If you spill liquid on the equipment, battery, or accessories, or they are accidentally
immersed in liquid, contact your service personnel or EDAN service engineer.
NOTE:
Automated cleaning/disinfection to the equipment and accessories is prohibited.

Cleaning

6.2.3
If the equipment or accessory has been in contact with the patient, then cleaning and
disinfection is required after each use. If there has been no patient contact and there is no
visible contamination then daily cleaning and disinfection is appropriate.
The validated cleaning agents are:
⚫ Ethanol (75%)
⚫ Isopropanol (70%)
Cleaning agents should be applied or removed using a clean, soft, non-abrasive cloth or
paper towel. Refer to the cleaning agent manufacturer's instruction for use with reference
to concentration, temperature and contact time.
Make sure that the power is off and the power cords are disconnected prior to
cleaning or disinfecting the device.
Any residue of cleaning solution should be removed from the main unit and
accessories after cleaning.
To clean the main unit:
1. Switch off the main unit and disconnect it from the power cord.
2. Wipe the exterior surface of the equipment using a soft cloth dampened with the
cleaning solution until no visible contaminants remain.
3. Wipe off the cleaning solution with a fresh cloth or paper towel dampened with tap
water after cleaning until no visible cleaning agent remains.
4. Dry the main unit in a ventilated and cool place.
To clean the sensor, probe and sensor cable:
1. Wipe them with a soft cloth dampened with the cleaning solution until no visible
contaminants remain.
2. Wipe off the cleaning solution with a fresh cloth or paper towel dampened with tap
water after cleaning until no visible cleaning agent remains.
3. Wipe off with a dry cloth to remove residual moisture.
CAUTION
WARNING
CAUTION
52
Care and Cleaning