PA Series Biofeedback and Stimulation System User Manual
Conducted RF
3 V
rms
IEC 61000-4-6
150 kHz to 80 MHz
6Vrms
between
0,15 MHz and 80
MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.7 GHz
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
3 V
rms
150 kHz to 80
MHz
c)
in ISM bands
6Vrms
bands
between
0,15 MHz and
80 MHz
3 V/m
80 MHz to 2.7
GHz
=
d
1
2 .
P
c)
in ISM
=
d
1
2 .
P
=
d
2
3 .
P
d =
6 P
/
communications
(Portable RF communications equipment
(including peripherals such as antenna
cables and external antennas) should be
used no closer than 30 cm (12 inches) to
any
part
Stimulation
specified by the manufacturer).
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
a)
survey,
compliance
b)
range.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
66
EMC Information
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.7 GHz
E
at
RF
equipment
of
the
Biofeedback
System,
including
should
be
less
level
in
each
wireless
bands
and
cables
than
the
frequency