Centurion Quantum comes in a variety of models with various configurations. Depending on the type of Centurion Quantum purchased, this manual may contain information that is not applicable to your model. This manual is divided into various sections and sub-sections to enable the user to access the required information with ease.
2 Description of Centurion Quantum 2.1 The Instrument Centurion Quantum is a wall mounted alcohol breath testing machine. It is designed for high volume, fixed point commercial and industrial alcohol breath testing applications. The Alcolizer Centurion Quantum incorporates innovative and patented replacement module technology, which means no more returning the Breathalyzer to base for service, Alcolizer Technology simply ships you an exchange module.
3.1.3 Choosing a suitable mounting location The Centurion Quantum is designed to be a rugged device however consideration must be made to its physical location and the possible environmental impacts on the instrument, particularly from the weather.
WARNING Do not use adhesives as the method of fixing as these are not considered to be a reliable or safe fixing means for the Centurion Quantum. 3.2 Mounting and Starting the Instrument 3.2.1 Mounting Mount the instrument in accordance with the following procedures: 1.
There may be a specific customer welcome screen. b. Take a straw from the straw holder provided. a. Alcolizer recommends the use of individually wrapped straws. c. Place the straw in the sample port CAUTION Never inject fluids into the sample port.
Do not use alcohol-based cleaning products, abrasive cleaning pads or chemicals. Non-alcohol-based cleaning wipes can be used to clean the Centurion Quantum As part of routine cleaning, use the brush provided to clean out the Centurion Quantum sample port as follows: ...
During the procedures ensure only the buttons mentioned are activated. Do not come in contact with any exposed components. Inside the Centurion Quantum are buttons which can be used to access certain menus during setup and fault finding. Users should not access the machine and press these buttons unless following instructions in this manual, or instructions verbally given when in contact with your Alcolizer Technology technical representative.
Centurion to access certain menus during setup and fault finding. Users should not access the machine and attach a keyboard unless specifically following instructions in this manual or verbal instructions when in contact with your Alcolizer Technology technical representative.
To access the User Menu, perform the following steps: 1. Connect a keyboard in accordance with paragraph 6.3. 2. Press the F1 key to access the Centurion Quantum User Menu area. 3. Press the left Alt key to bring up available menu choices.
Serial Number: A number that uniquely identifies the Centurion Quantum. This number is also printed on the side panel safety label located on the side of the instrument. Software Version: The version of software running in the Centurion Quantum (e.g., fm-17.0). 7.2.2 Alcohol Sample Module ...
Please contact your Alcolizer Technology representative for further advice. 7.8.5 Alcolizer Portal The Alcolizer Portal enables your device to post logs and status information from the instrument to the portal, which is a Cloud service. This portal provides a real-time solution with analytical and statistical tools.
Press Tab Right arrow to Services tab Press Tab twice to select Alcolizer Portal Press Space bar to insert X to enable Alcolizer Portal Wait for status to show as below Succeeded Status ...
To upgrade the software on your Centurion, you will need a USB stick. Follow instructions below to receive/download the latest software version. NOTE Contact Alcolizer to receive the latest software version or access our website for further information. 9.1 Saving software into the USB stick 1.
The Centurion instrument will require calibration of the Exchangeable Sample System every twelve (12) months. This service is to be completed by a trained Alcolizer Technology service technician, however due to our patented technology and design only the Alcohol Sample Module (ASM) is required to be exchanged during the calibration process.
10.3.3 Main Module servicing There is no requirement to service the main module, only the ASM, but on replacement of the ASM it is advisable to inspect the state of the chassis and screen and advise Alcolizer if there are any concerns.
Centurion Quantum User Manual 11 Printer Setup Alcolizer Centurion supports ONLY the following printers Epson TM-T82III Epson TM-T20 USB model Epson TM-T82IIIL To set up the Epson TM-T20 USB model: 1. Unpack printer. 2. Remove any tape used to secure parts for transport, then fit paper roll.
Web: www.alcolizer.com 12.1.3 Returning equipment to Alcolizer If the machine is to be returned to Alcolizer, prepare it for shipment in accordance with the following procedures: Unlock and open the instrument. Turn off the instrument by pressing & releasing the OFF button.