Alcolizer Centurion Quantum User Manual

For more information, please visit alcolizer.com
Table of Contents

Advertisement

Quick Links

Centurion Quantum
User Manual
The BrAC reading obtained by correct use of this device is only considered accurate at the
time of testing. Great care has been taken to ensure the accuracy of each reading.
Neither the manufacturer, the distributor, nor the owner accepts liability or responsibility
due to any action or claim arising from the reading produced by this device, whether used
REF: Centurion Quantum User Manual V2.docx
Uncontrolled when printed.
correctly or incorrectly.
Page 1
© Alcolizer Technology

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Centurion Quantum and is the answer not in the manual?

Questions and answers

Summary of Contents for Alcolizer Centurion Quantum

  • Page 1 Neither the manufacturer, the distributor, nor the owner accepts liability or responsibility due to any action or claim arising from the reading produced by this device, whether used correctly or incorrectly. REF: Centurion Quantum User Manual V2.docx Page 1 © Alcolizer Technology...
  • Page 2: Table Of Contents

    Centurion Quantum User Manual Contents Introduction ........................5 Description of Centurion Quantum ................... 6 The Instrument ......................6 Accessories ........................6 Specifications....................... 6 Equipment Overview ....................7 2.4.1 External View ....................... 7 2.4.2 Internal View ......................7 Installation Procedure ......................9 Installing the Back Plate ....................
  • Page 3 Email Status ......................31 7.9.6 Email Reports ..................... 31 7.9.7 Breath Test Results .................... 31 7.9.8 Instrument Status ....................32 7.9.9 Event Log Report ....................32 REF: Centurion Quantum User Manual V2.docx Page 3 © Alcolizer Technology Uncontrolled when printed.
  • Page 4 12 Troubleshooting ....................... 42 12.1 Contacting Service ..................... 42 12.1.1 Routine servicing of the ASM ................42 12.1.2 Contacting Alcolizer ................... 42 12.1.3 Returning equipment to Alcolizer ..............42 12.2 Troubleshooting FAQ ....................43 REF: Centurion Quantum User Manual V2.docx Page 4 ©...
  • Page 5: Introduction

    Centurion Quantum comes in a variety of models with various configurations. Depending on the type of Centurion Quantum purchased, this manual may contain information that is not applicable to your model. This manual is divided into various sections and sub-sections to enable the user to access the required information with ease.
  • Page 6: Description Of Centurion Quantum

    2 Description of Centurion Quantum 2.1 The Instrument Centurion Quantum is a wall mounted alcohol breath testing machine. It is designed for high volume, fixed point commercial and industrial alcohol breath testing applications. The Alcolizer Centurion Quantum incorporates innovative and patented replacement module technology, which means no more returning the Breathalyzer to base for service, Alcolizer Technology simply ships you an exchange module.
  • Page 7: Equipment Overview

    Weighing under 10kg, Centurion can be wall mounted in multiple, accessible locations. A large display screen also makes for clearer and easier reading of displayed results. 2.4.1 External View 2.4.2 Internal View REF: Centurion Quantum User Manual V2.docx Page 7 © Alcolizer Technology Uncontrolled when printed.
  • Page 8 Centurion Quantum User Manual Mounting Holes Male Lift Off Hinge Mounting Holes REF: Centurion Quantum User Manual V2.docx Page 8 © Alcolizer Technology Uncontrolled when printed.
  • Page 9: Installation Procedure

    3.1.3 Choosing a suitable mounting location The Centurion Quantum is designed to be a rugged device however consideration must be made to its physical location and the possible environmental impacts on the instrument, particularly from the weather.
  • Page 10: Mounting And Starting The Instrument

    WARNING Do not use adhesives as the method of fixing as these are not considered to be a reliable or safe fixing means for the Centurion Quantum. 3.2 Mounting and Starting the Instrument 3.2.1 Mounting Mount the instrument in accordance with the following procedures: 1.
  • Page 11: Initial Start-Up

    2. The instrument will take several minutes to start up. The loading screen will be visible. 3. The instrument is ready to use when the display screen is visible and ready to receive a sample. REF: Centurion Quantum User Manual V2.docx Page 11 © Alcolizer Technology Uncontrolled when printed.
  • Page 12: Operating Instructions

    There may be a specific customer welcome screen. b. Take a straw from the straw holder provided. a. Alcolizer recommends the use of individually wrapped straws. c. Place the straw in the sample port CAUTION Never inject fluids into the sample port.
  • Page 13: Analysing The Breath Sample

    The colour of the display is programmed to change according to the level of alcohol detected:  If the BrAC detected is zero, the display will be green REF: Centurion Quantum User Manual V2.docx Page 13 © Alcolizer Technology Uncontrolled when printed.
  • Page 14  doesn’t blow for long enough. If the blow is invalid, cease blowing, wait until the Insert Straw and Blow screen is displayed and then provide another sample. REF: Centurion Quantum User Manual V2.docx Page 14 © Alcolizer Technology Uncontrolled when printed.
  • Page 15: Owner Servicing

    Do not use alcohol-based cleaning products, abrasive cleaning pads or chemicals. Non-alcohol-based cleaning wipes can be used to clean the Centurion Quantum As part of routine cleaning, use the brush provided to clean out the Centurion Quantum sample port as follows: ...
  • Page 16: User Controls

    During the procedures ensure only the buttons mentioned are activated. Do not come in contact with any exposed components. Inside the Centurion Quantum are buttons which can be used to access certain menus during setup and fault finding. Users should not access the machine and press these buttons unless following instructions in this manual, or instructions verbally given when in contact with your Alcolizer Technology technical representative.
  • Page 17: Status Led Lights

    Centurion to access certain menus during setup and fault finding. Users should not access the machine and attach a keyboard unless specifically following instructions in this manual or verbal instructions when in contact with your Alcolizer Technology technical representative.
  • Page 18 Use only the left Alt key. The right Alt key has no function. The spacebar has the same functionality as the Enter key in normal usage. When entering data, use the spacebar not the Enter key. REF: Centurion Quantum User Manual V2.docx Page 18 © Alcolizer Technology...
  • Page 19: User Interface

    To access the User Menu, perform the following steps: 1. Connect a keyboard in accordance with paragraph 6.3. 2. Press the F1 key to access the Centurion Quantum User Menu area. 3. Press the left Alt key to bring up available menu choices.
  • Page 20: Instrument

     Serial Number: A number that uniquely identifies the Centurion Quantum. This number is also printed on the side panel safety label located on the side of the instrument.  Software Version: The version of software running in the Centurion Quantum (e.g., fm-17.0). 7.2.2 Alcohol Sample Module ...
  • Page 21: About

     Expiry Time: Once a single breath test is authorised, the test analysis must be completed within this time, or the authorisation expires, and a new authorisation is required. REF: Centurion Quantum User Manual V2.docx Page 21 © Alcolizer Technology...
  • Page 22: Shutdown Instrument

    From the User Menu, access the custom appearance screen by scrolling down to Custom Appearance and pressing the “Enter” key. The Custom Appearance screen will appear. 3. Tick the Show Idle Background Image box. REF: Centurion Quantum User Manual V2.docx Page 22 © Alcolizer Technology Uncontrolled when printed.
  • Page 23: Custom Text

    7.6 Intervals It is possible to adjust the timings for some screen display intervals. Descriptions of how the timings apply to the display are presented for each adjustable interval. REF: Centurion Quantum User Manual V2.docx Page 23 © Alcolizer Technology...
  • Page 24: Logs

     Notice Cleared. A notice status item cleared.  Notice Set. A notice status item set.  Breath Analysis Block Set. Data analysis is disabled.  Breath Analysis Block Cleared. Data analysis enabled. REF: Centurion Quantum User Manual V2.docx Page 24 © Alcolizer Technology Uncontrolled when printed.
  • Page 25: Export

    7.7.4 Erase Event Logs To erase the current event log records on the device, go to Logs / Export menu and use the Erase Event Log button. REF: Centurion Quantum User Manual V2.docx Page 25 © Alcolizer Technology Uncontrolled when printed.
  • Page 26: Networks

    4. Complete the Configuration page with the required details to establish the network connection if a Static IP address will be used. Alternatively select “Obtain an IP address automatically”. REF: Centurion Quantum User Manual V2.docx Page 26 © Alcolizer Technology...
  • Page 27: Services

    Please contact your Alcolizer Technology representative for further advice. 7.8.5 Alcolizer Portal The Alcolizer Portal enables your device to post logs and status information from the instrument to the portal, which is a Cloud service. This portal provides a real-time solution with analytical and statistical tools.
  • Page 28: Enabling Alcoconnect Data Management Portal

     Press Tab  Right arrow to Services tab  Press Tab twice to select Alcolizer Portal  Press Space bar to insert X to enable Alcolizer Portal  Wait for status to show as below  Succeeded Status ...
  • Page 29: Email

     Email Footer: Text to be appended to the body of emails produced by the instrument. It is also possible to use text replacement allowing real data to be inserted into the text when the email is sent. REF: Centurion Quantum User Manual V2.docx Page 29 © Alcolizer Technology...
  • Page 30: Settings - Server Test Mail

    Mail Server settings, it can take a very long time for the test email to fail, and this can also block further test emails being sent until the time out for the previous email has sent. REF: Centurion Quantum User Manual V2.docx Page 30 © Alcolizer Technology...
  • Page 31: Email Status

    If instrument authentication is enabled, the report can also include information about previous breath tests from the same authentication code. REF: Centurion Quantum User Manual V2.docx Page 31 © Alcolizer Technology Uncontrolled when printed.
  • Page 32: Instrument Status

    The Centurion is equipped with three relay outputs on the circuit board and their outputs can be configured to trigger in response to a breath test result. If a relay output is enabled, the relay will trigger for the specified duration. REF: Centurion Quantum User Manual V2.docx Page 32 © Alcolizer Technology...
  • Page 33: Hardware

    The Relay Status indicates if the relay is currently ON or OFF. There are three buttons on this page that briefly activate the relays. (Located under TEST) REF: Centurion Quantum User Manual V2.docx Page 33 © Alcolizer Technology Uncontrolled when printed.
  • Page 34: Functions

    Note: If the log data to download is small, you may not see the progress bar when back on the operational screen. The download time will differ depending on log data size. REF: Centurion Quantum User Manual V2.docx Page 34 © Alcolizer Technology...
  • Page 35: Self-Check

    One, many, or all tests can be run. Some tests require a degree of user interaction and if this occurs instructions will appear on the screen indicating how to proceed. REF: Centurion Quantum User Manual V2.docx Page 35 © Alcolizer Technology...
  • Page 36: Breath Test Authorisation

    The Americium serial protocol operates on both serial ports with the following settings:  Protocol version: am-spec-02  Date/Time: local time (not UTC) Please contact your Alcolizer Technology representative for further advice. REF: Centurion Quantum User Manual V2.docx Page 36 ©...
  • Page 37: Upgrade Software

    To upgrade the software on your Centurion, you will need a USB stick. Follow instructions below to receive/download the latest software version. NOTE Contact Alcolizer to receive the latest software version or access our website for further information. 9.1 Saving software into the USB stick 1.
  • Page 38: Checking The Software Version In Your Centurion

    Once centurion has restarted and is displaying the normal idle screen confirm the installed firmware version in the configuration status page. To access the status page, press FN1 switch on the back of the unit. REF: Centurion Quantum User Manual V2.docx Page 38 © Alcolizer Technology...
  • Page 39: Calibration

    The Centurion instrument will require calibration of the Exchangeable Sample System every twelve (12) months. This service is to be completed by a trained Alcolizer Technology service technician, however due to our patented technology and design only the Alcohol Sample Module (ASM) is required to be exchanged during the calibration process.
  • Page 40: Exchanging The Asm

    10.3.3 Main Module servicing There is no requirement to service the main module, only the ASM, but on replacement of the ASM it is advisable to inspect the state of the chassis and screen and advise Alcolizer if there are any concerns.
  • Page 41: Printer Setup

    Centurion Quantum User Manual 11 Printer Setup Alcolizer Centurion supports ONLY the following printers  Epson TM-T82III  Epson TM-T20 USB model  Epson TM-T82IIIL To set up the Epson TM-T20 USB model: 1. Unpack printer. 2. Remove any tape used to secure parts for transport, then fit paper roll.
  • Page 42: Troubleshooting

     Web: www.alcolizer.com 12.1.3 Returning equipment to Alcolizer If the machine is to be returned to Alcolizer, prepare it for shipment in accordance with the following procedures:  Unlock and open the instrument.  Turn off the instrument by pressing & releasing the OFF button.
  • Page 43: Troubleshooting Faq

    Life has Expired”. requires its service. Contact Alcolizer immediately to arrange a replacement module and have your serial number ready to quote. T: 1300 789 908 service@alcolizer.com www.alcolizer.com REF: Centurion Quantum User Manual V2.docx Page 43 © Alcolizer Technology Uncontrolled when printed.

Table of Contents